Instructions for use
|
Atlan A300, A300 XL, A350, A350 XL SW 1.0n
159
Alarms
10.4
Adopting alarm settings when changing the ventilation
mode
When the ventilation mode is changed, the alarm settings are also adapted.
Depending on the mode, alarm settings can either be adopted or set to Off.
For some modes, it can be configured whether or not the settings are adopted.
Observe the information in the following section: "Vertical tab "Config. 2"", page 171
However, the settings can be adjusted during operation at any time.
Alarm or alarm limit
CBM mode
Pause
Man/Spon,
PSV with
∆
Psupp
<5 hPa (cmH
2
O)
Ext. FGO
VC,
VC - AF,
PC,
PSV with
∆
Psupp
≥
5 hPa
(cmH
2
O)
FiO
2
low
1)
1) With inspiratory O
2
measurement with O
2
sensor
Is adopted.
Is adopted.
Not measured
Is restored or remains active.
FiO
2
Configurable
Configurable
FiO
2
low
2)
2) With integrated patient-gas measurement module
Is adopted.
Apnea (no CO
2
)
Off
Off
3)
3) The alarm remains suppressed until respiratory activity is detected.
Is adopted.
etCO
2
high
Configurable
etCO
2
inCO
2
FiO
2
Configurable
inAgent low
xMAC low
Not activated
Paw high
Is adopted.
Is adopted.
Not measured
Paw low
MV high
Configurable
Off
Configurable
MV low
Apnea (no flow)
Configu-
rable
4)
4) This alarm is only activated when the MV low alarm limit is also activated.
Off3)
On
Apnea (no pressure)
Off
Off