Maintenance

Maintenance

Calibration

6 monthly, only by authorized persons. An extension of the calibration interval, e.g. to 1  year, 
is possible. Prerequisite: analysis of the sensor data and further quality assurance mea-
sures by Dräger Safety Service.

Preliminary measures

Access 

» Calibration «

 in the main menu and select the menu item 

» Options «

In the menu item 

» Options «

, set the 

» Set point «

, the

 » Gas type «

 and, with the selection 

of

 "dry gas"

 and non-equipped ambient pressure sensor, the 

» Ambient pressure «

.

NOTE

The setting of the gas type is 

only

 for the calibration, the setting of the ambient pressure is 

only

 important for the dry gas calibration and has no bearing on the measurement of the 

subject.

Summary of Contents for Alcotest 7510

Page 1: ...D de Atemalkohol Messger t Technisches Handbuch Seite 2 bis 28 en Breath Alcohol Monitor Technical Handbook page 30 to 56 00123997_02 eps Dr ger Alcotest 7510...

Page 2: ...10 bersicht 10 Das Hauptmen 11 Men Information 11 bersicht 11 Letzte Tests anz 11 Abs Luftdruck anz 12 Kalibrierdaten anz 12 Kalibriertest Daten 12 Konfiguration anz 12 Testnummer anzeigen 12 Ger tein...

Page 3: ...3 Inhalt Datenschnittstelle 25 Ergebnisprotokoll ausdrucken 26 Ausdruck von gespeicherten Testergebnissen 26 Konfiguration des Ergebnisprotokolls 27 Stichwortverzeichnis 28...

Page 4: ...Diese Signalworte und die zugeh rigen Gefahren lauten wie folgt Alcotest ist eine eingetragene Marke von Dr ger VORSICHT Vor dem Ver ndern von Ger teparametern ist sicherzustellen dass die Anweisungen...

Page 5: ...r ck Men struktur Neben dem Messbetrieb gibt es weitere Bedienfunktionen im Men betrieb Der Men betrieb gliedert sich in zwei Ebenen Kurzmen und Hauptmen Kurzmen Das Kurzmen ist f r jeden Bediener anw...

Page 6: ...glicht die Einstellung von individuellen Ger teparametern Men Kalibrierung Dieses Men erm glicht das Kalibrieren des Ger tes Der PIN f r das Hauptmen kann ge ndert werden Seite 16 Einstellung des PIN...

Page 7: ...ests anz Seite 11 Kalibrierung Seite 19 Abs Luftdruck anz Seite 12 Kalibrierdaten anz Seite 12 Kalibriertest Daten Seite 12 Kon guration anz Seite 12 Testnummer anzeigen Seite 12 Ger teinformationen S...

Page 8: ...s wird aufgerufen Taste dr cken Die Eingabeanzeige erscheint Mit den und Tasten nacheinander die vier Stellen des PIN s einstellen und jeweils mit Taste best tigen Nach der vierten Eingabe springt de...

Page 9: ...n und Untermen s aufge f hrt Funktion bei Aktivierung k nnen Funktionen in ein oder mehreren Arbeitsschritten ausgef hrt werden Auswahl aktiviert bei Funktionen die ausgew hlt und aktiviert werden k n...

Page 10: ...zmen s kann mit der PC Software Dr ger Diagnostics siehe Daten schnittstelle auf Seite 25 erweitert werden z B Die einzelnen Funktionen des Kurzmen s sind in der Gebrauchsanweisung Dr ger Alcotest 751...

Page 11: ...em Dr ger Mobile Printer ausgedruckt werden siehe Gebrauchsanweisung Ergebnisprotokoll ausdrucken VORSICHT Vor dem Ver ndern von Ger teparametern ist sicherzustellen dass die Anweisungen dieses Handbu...

Page 12: ...siehe Seite 14 ver ndert werden Hier finden sich weitere Informationen zu diesen Punkten Anzeigemodus Darstellung des Endergebnisses in Ziffern oder als Bereich Messeinheit Messeinheit der Alkoholkonz...

Page 13: ...13 Das Hauptmen Men Information Ger teinformationen Anzeige von Sachnummer und Seriennummer Zur ck mit der Taste Softwareversion Anzeige der Version der Ger tesoftware Zur ck mit der Taste OK OK...

Page 14: ...Information Einstellungen Messger t Sprache w hlen Seite 14 Kalibrierung Messung Testnummer anzeigen Seite 15 Kontrast Hintergrund Seite 15 Datum einstellen Seite 15 Uhrzeit einstellen Seite 15 GPS E...

Page 15: ...s dem Ger t ausgelesen werden siehe Datenschnittstel le auf Seite 25 und ber eine Exportfunktion PC Programmen z B MS Excel oder anderen Visualisierungsprogrammen zur Verf gung gestellt werden In der...

Page 16: ...tenspeichers Statistikspeicher L schung bzw Zur cksetzen der Statistikdaten Diese beinhalten zum Beispiel Z hler f r die Anzahl der gemessenen Testergebnisse in den konfigurierten Messergebnisklassen...

Page 17: ...a Festes Volumen Probennahme erfolgt bei Erreichen des Minimalvolumens1 und Minimalblasdauer1 1 Standardeinstellung siehe Kriterien Atemprobe auf Seite 18 oder b Atemende Probennahme erfolgt nach Erre...

Page 18: ...tellung Standardeinstellungen Kriterien Atemprobe Einstellen des minimalen Atemvolumens und der minimalen Blaszeit f r eine g ltige Atem probe Nach Auswahl der Funktion Eingabe in Liter L mit den und...

Page 19: ...des bekannten Umgebungsdruckes oder die bekannte H he ber NN H he ber Meeresspiegel w hlbare Einheit Meter oder Fu mittels und Tasten eingegeben werden Der jeweils korrespondierende Wert wird angezeig...

Page 20: ...ndernfalls erscheint die Fehlermeldung in der Anzeige Kalibrierfehler In diesem Fall die Kalibrierung wiederholen Kalibriertest Nach Auswahl des Kalibriertest mit der Taste erscheint in der Anzeige da...

Page 21: ...r ger Safety Vorbereitende Ma nahmen Im Hauptmen Kalibrierung aufrufen und den Men punkt Optionen w hlen Im Men punkt Optionen den Sollwert den Gastyp und bei Auswahl Trocken gas und nicht best cktem...

Page 22: ...t ck Classic Abbildung I Adapter direkt in die Mundst ckaufnahme einsetzen Zur Vermeidung von Kondensation muss die Verbindung 3 so kurz wie m glich sein 5 cm 1 Muffe Bestell Nr 68 09 263 2 Adapter mi...

Page 23: ...insetzen Dr ger Alcotest 7510 ber einen Schlauch maximale L nge 1 m mit Pr fgasflasche ver binden 1 Muffe Bestell Nr 68 09 263 2 Adapter mit Wasserfalle Bestell Nr 68 05 473 3 Schlauch Viton 5 x 1 5 B...

Page 24: ...Kalibrierung ist ein Kalibrier Test ber den Men punkt Kalibriertest durchzuf hren Seite 20 Die Einstellungen Gastyp bleibt wirksam Die Einstellungen Umgebungsdruck bleibt wirksam sofern das Ger t nic...

Page 25: ...18 657 mit einem PC mittels PC Programm Dr ger Diagnostics 83 19 342 ab Version 1 60 W hrend der Kommunikation auf einer Schnittstelle ist die jeweils andere inaktiv PC Kommunikation Das Datenkabel z...

Page 26: ...uenz Wenn der Ausdruck beendet ist Men verlassen oder ggf weiteren Test selektieren Ablauf Ausdruck zum Abschluss einer Mess sequenz Dr ger Mobile Printer einschalten entsprechend der Gebrauchsanweisu...

Page 27: ...r ger Diagnostics 83 19 342 ab Version 1 60 gem ss der Anforderungen kon guriert werden Beispielausdruck Dr ger Mobile Printer Pr fort Alcotest 7510 Ger tenummer ARJS 0000 Drucker Nr ARXF 0017 Proben...

Page 28: ...25 Instandhaltung 21 Kalibrierung mit Nassgas 22 Kalibrierung mit Trockengas 23 Kurzmen 5 8 10 Mark II Simulator 22 Men Einstellungen 6 14 Men Information 5 11 Men Kalibrierung 6 19 Men betrieb 8 Men...

Page 29: ...29...

Page 30: ...the menu 37 The quick menu 38 Overview 38 The main menu 39 Information Menu 39 Overview 39 Show last tests 39 Show abs pressure 40 Show cal dates 40 Show acc dates 40 Show configuration 40 Show test...

Page 31: ...31 Contents Data interface 53 Print the result protocol 54 Printout of saved test results 54 Configuration of the result protocol 55 Index 56...

Page 32: ...h will alert you to the degree of hazard you may encounter These signal words and corresponding hazards are as follows Alcotest is a registered trademark of Dr ger CAUTION Before changing the device p...

Page 33: ...ure In addition to the measuring mode further operational functions are available in the menu mode The menu mode is divided into two levels Quick menu and Main menu Quick menu The quick menu can be se...

Page 34: ...menu enables the setting of individual device parameters Calibration Menu This menu enables the calibration of the device The PIN for the main menu can be changed page 44 Setting of the PINs on deliv...

Page 35: ...e 42 Show last tests page 39 Calibration page 47 Show abs pressure page 40 Show cal dates page 40 Show acc dates page 40 Show con guration page 40 Show test number page 40 Device information page 41 S...

Page 36: ...tion for the PIN entry is selected Press the key The entry screen appears With the and keys set the four digits of the PIN and con firm each one with the key After the fourth entry the cursor jumps to...

Page 37: ...here are listed under this item Function If activated functions can be executed in a single step or several steps Selection activated For functions which can be selected and activated activation is se...

Page 38: ...the quick menu can be expanded with the PC software Dr ger Diagnostics see Data interface on page 53 e g The individual functions of the quick menu are described in detail in the Dr ger Alcotest 7510...

Page 39: ...n be printed on a Dr ger Mobile Printer see Instructions for Use print result protocol CAUTION Before changing the device parameters the content of the instructions in this manual must be completely u...

Page 40: ...menu see page 42 Further information about these items can be found here Display mode Display of the net results in digits or as a range Measurement unit Measurement unit of the alcohol concentration...

Page 41: ...The main menu Information Menu Device information Display of the part number and serial number Go back with the key Software version Display the version of the device software Go back with the key OK...

Page 42: ...d the printout Information Settings Instrument Set language page 42 Calibration Measurement Show test number page 43 Set contrast backlight page 43 Set date page 43 Set time page 43 GPS settings page...

Page 43: ...83 19 342 starting with version 1 60 see Data interface on page 53 and made available using the export function of PC programs e g MS Excel or other visualiza tion programs During the test set up time...

Page 44: ...ion of the entire data memory Statistics Log Deletion or resetting of the statistics data This includes for example the counter for the number of measured test results in the con figured measurement r...

Page 45: ...the type of sample mode a Fixed volume Sampling occurs when the 1 minimum volume and minimum blowing time1 have been reached 1 Default setting see Blow criteria on page 46 or b End of breath Sampling...

Page 46: ...an example Default settings Blow criteria Setting of the minimum breath volume and the minimum blow time for a valid breath sample After selection of the function entry in liters L with the and keys...

Page 47: ...ting of the established ambient pressure or the established height above NN height above sea level selectable unit meter or feet can be entered using the and keys The corresponding value in each case...

Page 48: ...s not successful the error message Calibration error appears in the display In this case repeat the calibration Accuracy check After selection of the Accuracy check with the key the calibration gas to...

Page 49: ...rvice Preliminary measures Access Calibration in the main menu and select the menu item Options In the menu item Options set the Set point the Gas type and with the selection of dry gas and non equipp...

Page 50: ...sic mouthpiece Figure I Insert the adapter directly into the mouthpiece receptacle The connection 3 must be as short as possible 5 cm in order to prevent condensation 1 Coupling Order No 68 09 263 2 A...

Page 51: ...ptacle Connect the Dr ger Alcotest 7510 to the calibration gas bottle using a hose maximum length 1 m 1 Coupling Order No 68 09 263 2 Adapter with water trap Order No 68 05 473 3 Hose Viton 5 x 1 5 Or...

Page 52: ...sful calibration an accuracy check must be conducted using the menu item Accuracy check page 48 The Gas type settings remain effective The Ambient pressure settings remain effective unless the device...

Page 53: ...18 657 with a PC using the PC program Dr ger Diagnostics 83 19 342 starting with version 1 60 During the communication with an interface the other interface is inactive PC communication Connect the d...

Page 54: ...the menu or select an additional test as applicable Printout procedure upon completion of a measuring sequence Switch on the Dr ger Mobile Printer as per the Dr ger Mobile Printer Instructions for Us...

Page 55: ...quirements using the PC pro gram Dr ger Diagnostics 83 19 342 starting with version 1 60 Sample printout Dr ger Mobile Printer Test location Alcotest 7510 Device number ARJS 0000 Printer No ARXF 0017...

Page 56: ...n 51 Enter PIN 36 Ethanol calibration gas bottle 51 Ethanol standard solution 50 Infrared interface 53 Main menu 33 35 36 39 Main menu overview 35 Maintenance 49 Mark II Simulator 50 Menu functions 33...

Page 57: ...57...

Page 58: ......

Page 59: ......

Page 60: ...n 01 January 2008 Subject to alteration Dr ger Safety AG Co KGaA Revalstra e 1 D 23560 L beck Germany Tel 49 451 8 82 0 Fax 49 451 8 82 20 80 www draeger com Importer UK Draeger Safety UK Ltd Ullswate...

Reviews: