Dräger Alcotest 7410 GLC Operator'S Manual Download Page 6

Alcotest 7410 GLC User’s Manual

4

Operation of The Alcotest 7410 GLC

Before using the GLC, the operator should

be aware of the calibration and battery

charge status.

A. Calibration Status

The calibration of the GLC is checked every

two weeks to ensure proper results are

obtained. This duty can only be performed

by a designated 

calibrator 

who has taken

the required training. Upon completing the

calibration process, the calibrator affixes a

sticker showing the date of the last

calibration. If the current date is more than

two weeks from the last calibration date, do

not use the GLC. Return the device to a

calibrator.

B. Battery Charge

The power supply for the GLC is either a

rechargeable nickel cadmium (NiCd) battery

pack or a non-rechargeable alkaline battery

pack containing three disposable C-size 1.5

volt batteries. Regardless of what power

supply is used, the device has a 

“Lo Bat”

indicator which will appear in the lower left

hand corner of the liquid crystal display

(LCD). The device must be turned on to see

this indicator. If you are using NiCd

batteries, a “Lo Bat” indicator means battery

power is low and operation can be

continued for only a few minutes. Return

the device to the charging unit. If you are

using disposable alkaline batteries, “Lo Bat”

will come on as part of the normal

operational cycle. It does not indicate that

the batteries must be replaced. When there

is insufficient power to conduct a test, the

instrument will automatically shut off.

NOTE: The users must be aware of which

device they have. Only GLCs which have

rechargeable batteries should have the

charging unit at your detachment. The type

of batteries your device has should be

confirmed with the calibrator. This is

important not only for operational reasons,

but for safety reasons as well. Do not

attempt to charge a GLC that has

non-rechargeable batteries.

C. Operational Procedure

Once you have determined the calibration

and battery charge status, you are ready to

proceed into the normal operational cycle.

This cycle consists of the following:

1. Press the button (shaded grey) on the

right side of the device 

once.

2. The 

yellow

 “On” light illuminates and a

self test is carried out.

3. A brief indication of the display option

for the device is shown, followed by the

model designation (GLC).

4. This is followed by the display

“ - - - wait ”

5. After 10 to 30 seconds the 

“wait”

indication goes out and the green

Summary of Contents for Alcotest 7410 GLC

Page 1: ...Alcotest 7410 GLC Operator Manual Prepared By Toxicology Services National Forensic Services Amended April 2011 ...

Page 2: ...s And Maintenance 9 Law and Field Use of the Alcotest 7410 GLC 10 Glossary and Abbreviations 13 Note The instructions procedures and policies included in this manual were in effect at the time of its writing They are subject to revisions imposed by amendments to the Criminal Code Saskatchewan Traffic Safety Act or the Vehicle Administration Act case law decisions manufacturers instructions and cha...

Page 3: ...ivers are apprehended in this manner and not impaired drivers Intoxicated drivers are severely impaired and demonstrate gross physical symptoms Impaired drivers show few symptoms but still represent a high risk of being responsible for a motor vehicle accident To detect impaired drivers the peace officer needs to use the second means of acquiring reasonable and probable grounds This is a roadside ...

Page 4: ...ding the lungs brain and liver It is the depressant action of alcohol in the brain that causes impairment and intoxication Elimination of the alcohol begins immediately after it has entered the blood Most of the alcohol 90 98 is eliminated by metabolism in the liver The remainder 2 10 is eliminated unchanged through urine sweat breath and other body fluids Unlike absorption the elimination process...

Page 5: ...Alcotest 7410 GLC User s Manual 3 Description ...

Page 6: ... battery power is low and operation can be continued for only a few minutes Return the device to the charging unit If you are using disposable alkaline batteries Lo Bat will come on as part of the normal operational cycle It does not indicate that the batteries must be replaced When there is insufficient power to conduct a test the instrument will automatically shut off NOTE The users must be awar...

Page 7: ...displayed Note and record the result 10 Turn off the device by pressing the side button twice D Improper Sample If the sample was interrupted too short or too weak an E0 indication is briefly displayed The Ready light goes out and the tone ceases followed by two short audible signals When the Ready light reappears instruct the person to blow again No results will be displayed until a proper breath...

Page 8: ... A microprocessor evaluates this current to determine the BAC The BAC is displayed in the LCD window Normally the result is provided in 10 to 25 seconds As indicated there are two reaction processes The first reaction is fast and allows the BAC to be measured quickly The second reaction is slow and must go to completion before another test can be conducted The time interval before the next test ca...

Page 9: ...ossibility of a falsely high ASD test NOTE High alcohol concentrations obtained from mouth alcohol could shorten the life of the electro chemical sensor 2 Smoking Tobacco smoke in the expired air can cause damage to the electro chemical sensor If the subject taking the test is smoking it is important that you have the person stop Then wait for at least three minutes before taking a breath sample 3...

Page 10: ...hol was discussed earlier If the observation period was not properly observed there may be residual alcohol in the mouth at the time of the GLC test This will cause the ASD test to be higher than the evidentiary test 3 Tolerance of the instruments All scientific instruments have a distinct precision It is often called the tolerance and is expressed with a plus and minus value e g 10 mg No instrume...

Page 11: ...t after switching on and 4 minute interval Instrument malfunction Refer to calibrator for possible return to Dräger Service No continuous tone when blowing into the instrument Instrument not ready for measurement Wait until green Ready light is lit E0 Subject not blowing hard or evenly enough Remind subject how to blow E1 Sensor fault Switch device off and on If E1 re appears sensor must be replac...

Page 12: ...ng driving offences approved screening device means a device of a kind that is designed to ascertain the presence of alcohol in the blood of a person and that is approved for the purposes of this section by order of the Attorney General of Canada The key here is that before any screening device can be used it must be approved for use and appear on the Approved Screening Devices Order The Alcotest ...

Page 13: ...erations Besides the Criminal Code the ASD user must be aware of Division or departmental policy related to ASD use The Operational Manual and F Division policy should be consulted regularly for up to date policy Some of the highlights in policy are presented here Municipal police officers should consult their own department policy manuals For reference policy sections dealing with ASD use are Ope...

Page 14: ...llowing actions are normally taken 1 0 to 49 mg on the GLC Investigate possibility that drugs or illness are responsible for condition OR Release suspect unless grounds exists for suspension under graduated licensing 2 A ALERT suspend licence under Sec 146 1 146 3 T S A or Sec 78 2 V A A if applicable Release driver 3 F FAIL demand an evidentiary breath test in accordance with Section 254 3 a i C ...

Page 15: ...e e g Alcotest 7410 GLC BAC blood alcohol concentration C C Criminal Code evidentiary instrument an approved instrument pursuant to the Criminal Code e g Intoxilyzer 5000C or Intox EC IR II GLC Alcotest 7410 GLC T S A Traffic Safety Act V A A Vehicle Administration Act mg milligrams of alcohol in 100 millilitres of blood ...

Reviews: