Dräger Alcotest 6000 Instructions For Use Manual Download Page 5

Use

5

en

3.2

Intended use

The Dräger Alcotest

®

 6000

1)

 breath-alcohol analyzer is used to measure the 

alcohol concentration in human breath. The measurement should not be used as 
grounds for medical or therapeutic treatments.

4

Use

WARNING

Small parts of the device may become detached and be swallowed.
► The device is not intended for children under 3 years.

4.1

Prerequisites for use

The device is factory-calibrated when delivered. If the calibration date is longer 
ago than 12 months, the instrument should be calibrated, e.g., by DrägerService. 

Observe a wait time of approx. 15 minutes after the last consumption of 
alcohol.

The wait time cannot be reduced by rinsing the mouth with water or with 
non-alcoholic beverages.

Residual mouth alcohol or aromatic beverages (e.g., fruit juice), mouth 
sprays containing alcohol, medication and drops, as well as belching and 
vomiting may cause incorrect measuring values.

4.1.1

Before sampling

The subject must breathe regularly and calmly before sampling.

The subject must be able to provide the required minimum breath volume 
for the sampling.

4.1.2

Attaching the wrist strap

The wrist strap should be used to secure the device on the wrist.
1.

Guide the fastening lug on the wrist strap through the eye on the device.

2.

Pull the top part of the wrist strap through the fastening lug.

4.1.3

Preparation for use

1.

Ensure that the batteries are charged and inserted.

2.

Ensure that a clean mouthpiece is attached.

1)

Alcotest

®

 is a registered trademark of Dräger.

Summary of Contents for Alcotest 6000

Page 1: ...ow the Instructions for Use The user must fully understand and strictly observe the instructions Use the product only for the purposes specified in the Intended use section of this document enUS Instr...

Page 2: ...2 11 13 10 1 9 4 7 8 6 5 12 00100217 ai 2 3 14...

Page 3: ...ract for all maintenance activities and that all repairs are carried out by Dr ger Do not use a faulty or incomplete product Do not modify the product Notify Dr ger in the event of any component fault...

Page 4: ...asier 3 Description 3 1 Product overview see Page 2 1 Breath alcohol analyzer 2 button switch instrument on off confirm entry 3 button access menu set value 4 LEDs 5 Mouthpiece receptacle configurable...

Page 5: ...e reduced by rinsing the mouth with water or with non alcoholic beverages Residual mouth alcohol or aromatic beverages e g fruit juice mouth sprays containing alcohol medication and drops as well as b...

Page 6: ...e instrument will switch off automatically approx 2 minutes after the last use Switching off the instrument does not reduce the post test waiting period before you can perform another test 4 2 Initial...

Page 7: ...s or to change and press to confirm Measurement unit A cursor pre selects a measurement unit in the list Press the or button to select the needed measurement unit If the cursor selects the needed meas...

Page 8: ...tional self test is carried out The following is shown on the display for approx 4 seconds READY The instrument is ready to measure 4 Blow evenly and without interruptions into the mouthpiece A progre...

Page 9: ...nce of alcohol Carrying out passive sampling 1 Keep the button pressed for approx 1 second until the startup screen appears 2 Hold the instrument which is ready to measure into the air to be analyzed...

Page 10: ...o perform a test Hold the button for 2 seconds to leave the menu Bluetooth optional Menu option to switch Bluetooth on or off Select Bluetooth On or Off with the or button Confirm with the button The...

Page 11: ...ld get an accuracy check Hold the button for 2 seconds to leave the submenu Date time Date time as well as date format yyyy mm dd yyyy mm dd mm dd yyyy dd mm yyyy and time format 12h 24h can be adjust...

Page 12: ...ement unit The measurement unit for the alcohol test results can be adjusted within the following options g L mg 100 mL mg mg mL mg L g 100 mL g g L g 210 L Select the measuring unit using the and but...

Page 13: ...ing the instrument on and off several times contact Dr gerService or your dealer Fault message Cause Remedy The instrument cannot be switched on Battery drained Replace battery Volume too low sample i...

Page 14: ...brated with an adjustment service if an accuracy check does not produce a result per NHTSA s EBT Model Specifications Adjustment service should only be carried out by authorized persons e g Dr gerServ...

Page 15: ...liquid enters the gas inlet 2 After cleaning the device store it unpacked at room temperature 20 25 C for at least 24 hours 3 Afterward carry out passive sampling to ensure that no ethanol residue re...

Page 16: ...with applicable regulations Measuring principle Electrochemical sensor Measuring range Breath alcohol concentration 0 000 to 0 500 g 210L Blood alcohol concentration 0 000 to 0 500 BAC Small values w...

Page 17: ...g L 1 7 of the measured value Wait time for measurement readiness following a previous measurement with 0 25 mg L 20 s 0 50 mg L 40 s 1 00 mg L 80 s 1 50 mg L 120 s 0 050 20 s 0 100 40 s 0 200 80 s 0...

Page 18: ...tallation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the in...

Page 19: ...de mantenimiento sean realizados por Dr ger No utilizar productos defectuosos o incompletos No realizar modificaci n alguna en el producto Informar a Dr ger si se produjeran fallos o aver as en el pro...

Page 20: ...pogr ficas Este s mbolo identifica informaci n que facilita el uso del producto 3 Descripci n 3 1 Vista general del producto v ase p gina 2 1 Analizador de alcohol en el aliento 2 Bot n conectar desco...

Page 21: ...tivo deber calibrarse por ejemplo por parte de Dr gerService Esperar aprox 15 minutos despu s de haber tomado la ltima bebida alcoh lica El plazo de espera no puede reducirse enjuag ndose la boca con...

Page 22: ...mu eca 4 1 4 Encendido y apagado del instrumento Encender el instrumento 1 Mantener el bot n pulsada durante aprox 1 segundo 2 Esperar hasta que el instrumento indique que est listo para realizar la m...

Page 23: ...para seleccionar el formato de hora necesario Si el cursor selecciona el formato de hora necesario pulsar el bot n para confirmar Hora Un cursor preselecciona la hora del d a Pulsar el bot n o para ca...

Page 24: ...1 Retirar una boquilla n ref 68 10 690 convencional 68 11 055 con v lvula trasera de su embalaje Por motivos de higiene deje la envoltura protectora alrededor de la zona de contacto con la boca hasta...

Page 25: ...nueva medici n pulsar el bot n brevemente Tras el tiempo de recuperaci n del sensor v ase Datos t cnicos en la p gina 33 aparecer LISTO y el dispositivo est listo para usar de nuevo 4 4 Realizar una m...

Page 26: ...boquilla conforme a las normativas locales No reutilizar la boquilla 4 7 Control de men s Abrir el men Pulsar el bot n Navegar en el men o realizar ajustes Pulsar el bot n o Confirmar la funci n entra...

Page 27: ...n para ajustar la impresora est ndar y activa o desactiva la funci n de impresi n Seleccionar la impresora con el bot n o Confirmar con el bot n Resultados Resumen de las 10 ltimas pruebas realizadas...

Page 28: ...submen con la opci n Regresar Cambiar una configuraci n o un n mero establecido pulsando brevemente el bot n o p ej cambio de la hora Confirmar la selecci n manteniendo el bot n presionado durante 2...

Page 29: ...Unidad de medici n La unidad de medici n para los resultados de la prueba de alcohol se puede ajustar dentro de las siguientes opciones g L mg 100 mL mg mg mL mg L g 100 mL g g L g 210 L Cambiar la un...

Page 30: ...instrumento El n mero que aparece bajo el mensaje se utiliza para funciones de servicio Si el mensaje de estado se mantiene despu s de encender y apagar el instrumento varias veces contactar con Dr ge...

Page 31: ...en ingl s de la Administraci n nacional para la seguridad vial en EE UU El ajuste servicio solo deben realizarlo personas autorizadas p ej Dr gerService 6 1 Cambiar la pila 1 Apagar el aparato Rango...

Page 32: ...a El dispositivo se debe limpiar con un limpiador a base de etanol Limpiar el dispositivo 1 Frotar el dispositivo con un pa o de un solo uso impregnado con limpiador No aplicar el limpiador directamen...

Page 33: ...ucto a Dr ger sin coste alguno para su eliminaci n La informaci n a este respecto est disponible en las delegaciones nacionales y en Dr ger Las pilas y las bater as no deben eliminarse como residuos d...

Page 34: ...cenamiento De 20 a 60 C de 4 a 140 F De 15 a 75 r h de 10 a 100 r h 48 h Toma de muestras Volumen respiratorio 1 2 L preajustado Flujo respiratorio 10 L min preajustado Tiempo de soplado 1 6 s preajus...

Page 35: ...nt a alguna de que dichos ruidos de interferencia no se vayan a producir en una instalaci n concreta Si el presente aparato provocara ruidos de interferencia perjudiciales en la recepci n de radio o t...

Page 36: ...guientes medidas Reorientar o recolocar la antena receptora Aumentar la separaci n entre el aparato y el receptor Conectar el aparato a una salida de otro circuito distinto al que est conectado el rec...

Page 37: ...Co KGaA Revalstra e 1 23560 L beck Germany Tel 49 451 882 0 Fax 49 451 882 20 80 www draeger com 93 00 218 GA 4754 660 Dr ger Safety AG Co KGaA Edition 02 November 2021 Edition 01 October 2019 Subjec...

Reviews: