40
Faults, cause, remedy
Fault
Cause
Remedy
Dräger X-zone 5000 cannot
be loaded.
Charging unit plug is not in correct
contact with the Dräger X-zone
5000.
Ensure that the charging unit plug
is correctly plugged in.
Check the battery LED.
Battery service life is low.
Battery is not fully charged.
Load battery for at least 8 h - 12 Ah,
14 h - 24 Ah.
Battery not regularly charged.
Charge battery regularly even when
switched off, at least every
2 months.
Outdoor temperature is too low.
Warm up Dräger X-zone 5000.
Outdoor temperature is very high.
Use safety housing where
appropriate.
Battery is faulty.
Have the battery checked by
Dräger.
Inductive charging is not functioning
correctly as the distance between
the Dräger X-zone 5000 and the
charging station is too great.
Check the distance, remove any dirt
if necessary.
No wireless connection between the
Dräger X-zone 5000 devices.
Wireless function is not activated.
Activate the wireless function using
the Dräger CC-Vision PC software
(refer to “Configuring the device” on
page 39).
Dräger X-zone 5000 devices are
very far apart.
Position the Dräger X-zone 5000
devices more closely together.
Position additional devices in the
chain.
Place Dräger X-zone 5000 at
a higher position, use base (order
no. 83 20 645) where appropriate.
The wireless connection is made
more difficult by an industrial
environment: e. g. steel walls.
Position the Dräger X-zone 5000
devices more closely together.
Position additional devices in the
chain.
Connect the Dräger X-zone 5000
with a communication cable (refer to
“Connecting devices via a cable
connection” on page 34).
Dräger X-zone 5000 are covered
by conductive materials
(e. g. metal grids).
Ensure there are no obstructions.
Network number of the Dräger
X-zone 5000 devices is different.
Use Dräger X-zone 5000 devices
with the same network numbers.
The network number can be
configured using the Dräger CC-
Vision PC software (refer to
“Configuring the device” on
page 39).
Wireless frequency of the Dräger
X-zone 5000 devices is different.
Use Dräger X-zone 5000 devices
with the same wireless frequency.
Cable connection is not functioning. Cable plug is not correctly plugged
in, cable assignment is incorrect or
cable is broken.
Check cable connection and
assignment. Ensure that the cable
plug is correctly plugged in.
Alarm contact is not switching.
Cable plug is not correctly plugged
in, cable assignment is incorrect or
cable is broken.
Check cable connection and
assignment. Ensure that the cable
plug is correctly plugged in.