![Dräger 2M 20 116 Instructions For Use Manual Download Page 42](http://html1.mh-extra.com/html/drager/2m-20-116/2m-20-116_instructions-for-use-manual_2526506042.webp)
The system initially proposes the default temperature
setting 36.3
°
C.
1 The value 36.3
°
C flashes in the display.
2 Briefly press and release
p
or
P
to confirm the
value.
1 The displayed value remains constant.
The temperature setting can be adjusted in 0.1
°
C steps:
2 Hold down key
p
or
P
until the desired temperature
setting appears in the display.
3 The yellow LED
O
lights up or flashes according to
the adjustment (heater function display).
To check heater function during operation:
4 Press the
6
key.
5 The simulated temperature value must be 36
°
C
(
±
0.1
°
C) (function check).
If the sensor probe is disconnected or the sensor is
defective:
6 Three flashing dashes in the middle of the display
7 the red alarm LED marked Sensor flashes.
A continuous audible alarm tone is emitted.
●
Immediately connect the sensor plug or change the
skin temperature sensor.
If the error cannot be remedied immediately:
●
Remove the skin temperature sensor from the patient.
●
Switch to manual mode (»Man«).
See "Troubleshooting", page 50.
Operation
Using skin temperature control
42
3
2 2
1
4
5
7
6
Summary of Contents for 2M 20 116
Page 5: ...Montagem glichkeiten Montage an einem Fahrgestell Wandmontage 5 Zweckbestimmung 1 382 94...
Page 33: ...Installation options Mounted on wheeled stand Wall mounted 33 Intended Use 1 382 94...
Page 61: ...Possibilit s de montage montage sur un chariot montage mural 61 Domaine d application 1 382 94...
Page 89: ...Posibilidades de montaje Montaje sobre soporte m vil Montaje mural 89 Aplicaci n 1 382 94...
Page 114: ...114...
Page 115: ...115...