Dräger Medical AG & Co. KGaA
Service Mode Savina Version 3.0
Page 80
6HUYLFH0RGH
,QWHUQDO6HUYLFH0RGH
'HVFULSWLRQRIWKHWHVWVWHSV
7HVW6WHS
Test step 0 is used to read out the error log (error list). A list of error messages appears. 5 error
messages are visible at a time. Additional error messages can be viewed by scrolling. The error
messages have the following structure:
"Send Logbook" transmits the error log to a PC (for settings in the terminal program, see
also"
"). "Delete Logbook" deletes the error log.
For assignment of error numbers, see chapter "
7HVW6WHS
Test steps 1 and 2 are used to test the keys on the front panel. The name of the key appears, e.g.
"Nebulizer", and "Not press." (not pressed). "Pressed" is displayed as soon as the key is pressed.
This method can be used to test all keys.
7HVW6WHS
— see
First occurrence
Last occurrence
Error number
Number of errors