General
20 Remove the loudspeaker screws.
21 Remove the loudspeaker rom the faulty front frame and
place it aside.
22 Turn the front frame around.
23 Carefully remove the control knob from the rotary
transducer (Note: the control knob contains a loose leaf
spring).
24 Remove the nut and the washer (
/1).
Fig. 7
Rotary transducer
25 Remove the rotary transducer from the faulty front
frame and place it aside.
26 Remove the keypad labeling strip from the faulty front
frame and
place it aside
.
27 Dispose of the faulty front frame including front
membrane according to applicable waste disposal
regulations.
28 Attach the new “EvitaXL” label to the provided indented
space in the top right corner of the front frame.
29 Turn the new front frame around.
30 Secure the Adapter PCB to the new front frame using
the screws and washers.
31 Secure the loudspeaker to the new front frame using
the screws and washers.
32 Plug the loudspeaker connector onto the plug-in
contact of the Adapter PCB.
33 Secure the rotary transducer to the front frame using
the washer and nut.
34 Plug the rotary transducer connector onto the plug-in
contact of the Adapter PCB.
35 Push the control knob including leaf spring onto the
rotary transducer (Note: Observe clamping position of
the leaf spring, correct if necessary).
A faulty front frame including front membrane is
special waste. Dispose of a faulty front frame
including front membrane according to applicable
waste disposal regulations.