Dräger Medical AG & Co. KGaA • EvitaXL
1
Repair instructions
General
Front frame, complete
1
General
In case of repair, replace the “complete front frame”.
The “front frame, complete”, comprises the front frame,
the label (EvitaXL), and the front membrane.
1.1
Replacing the front frame
1 Switch off the device.
2 Unplug the power plug of the device from the socket-
outlet.
3 Disconnect the compressed-gas supply hoses from the
central gas supply.
4 Create ESD conditions.
5 Remove the control unit from the interlock of the device
and place it on a static-dissipative mat.
In case of repair, remove the keypad labeling strip
from the faulty front frame and use it in the new front
frame!
Electrostatic discharge may damage electrostatic
sensitive devices. When handling electrostatic
sensitive devices, use a static-dissipative mat
and a static dissipative wrist strap.