Dräger Medical AG & Co. KGaA • EvitaXL 

 3

Repair instructions

Service LEDs

Fig. 1

CPU PCB 68332

G2

Lithium battery

U1

DC/DC converter

D35

Real-time clock

D1

Microprocessor

D22

EEPROM

3

Service LEDs

The service LEDs indicate a reset.

LED Reset

Meaning

1

Overvoltage

+5 V is too high.

2

Undervoltage

+5 V is too low.

3

CPU reset

The microprocessor, which 
is located on the CPU 68332 
PCB, resets its own 
peripherals.

4

Reset up

The reset signal from the 
Pneumatics Controller PCB 
resets the CPU 68332 PCB.

5

Reset down

The reset signal from the 
CPU 68332 PCB resets the 
microprocessor on the 
Pneumatics Controller PCB 
and the boards on the 
motherboard.

6

Display reset

The reset signal from the 
CPU 68332 PCB resets the 
microprocessor of the 
control unit.