QUICK GUIDE BABYLOG® VN500
12 |
Getting started
Switching on Babylog VN500 and Infinity C500
Prerequisites:
– Babylog VN500, Infinity C500,
PS500 and GS500 are
reprocessed and assembled
ready for operation.
– The mains power supply and
the gas supply are connected.
– The Babylog VN500 toggle switch
is set to On
1. Press the key on Infinity C500.
The system is started. The start
dialog is displayed.
Babylog VN500 provides you with
two options:
– Use settings of previous patient
or
– Admit new patient
– If the Start dialog is closed using
the X button, Babylog VN500 adopts
the settings of the previous patient.
Ventilation unit,
left side:
Toggle switch
1.