Dr.Meter PH838 User Manual Download Page 3

Operation Instructions

Before Use

a) Open the battery cap and install the battery.

b) If the electrode has been changed, the instrument must be recalibrated.

c) We recommend recalibrating the unit once a month.

3 Point Calibration

In the measuring mode, press and hold the

button for 3s.

a) The LCD display will show a flashing [CAL 1] icon and the number 4.00

will appear. Rinse the electrode with distilled water and put it into a pH4.00

solution. When the reading is stable, the LCD display will display “PAS”,

which means the first calibration point has succeeded. Then the device will

proceed to the second calibration point --- the [CAL 2] icon will flash and the

number 6.86 will appear.

b) Rinse the electrode with distilled water and put it into the pH6.86 solution.

When the reading is stable, the LCD display will display “PAS”, which

means the second calibration point has succeeded. Then the device will

proceed to the third calibration point --- the [CAL 3] icon will flash and the

number 9.18 will appear.

c) Rinse the electrode with distilled water and put it into the pH9.18 solution.

When the reading is stable, the LCD display will read “PAS”, which means

the entire calibration has succeeded. Then the device will enter measuring

mode.

d) If the buffer value doesn’t match the preset value, you can correct the value

manually by pressing the

button or

button.

e) If [Err] appears in the LCD display, the calibration did not succeeed, which

means the standard buffer solution outranges the calibration point.

Timing