Mach LED 500DF
Dr. Mach
Lamps and Engineering
59190002
Edition 05
15.09.2014 / Bak We page
22/26
5.6 General remarks
When using more than one OT-lamp at the same time (OT-lamp combinations), due to the over-
lapping of the light fields of different lights, the total radiation intensity can exceed the value of
1000 W/m
2
. This means a risk of higher heat development in the wound field.
When using more than one OT-lamp at the same time (OT-lamp combinations), due to the light fields
overlapping of different lights the maximum permissible values for UV-radiation (< 400 nm) of 10 W/m²
can be exceeded.
The test certificate for the electrical safety test can be requested when needed. Please
provide the serial number of the respective light.
In case of a collective wiring of further lights or devices at installation, chapter 16 of the European
standard EN 60601-1:2006 must be applied and eventually it has to be checked if the requirements
are met.
The light must be tested according to EN 62353 at commissioning.
The polarity is very important for the installation of the light. In case the light does not
function after installation, the polarity must be checked at the secondary side of the power
supply.
6. CE-mark
The products Mach LED 500DF comply with the standards 93/42/EEC for medical
products of the European Community’s Council. Dr. Mach applies the standard
EN 60601-2-41.
Dr. Mach GmbH is certified according to DIN EN ISO 13485:2012 + AC:2012.
7. Disposal
The OT-lamp doesn’t contain any dangerous goods.
The components of the OT-lamp should be properly disposed at the end of its shelf-
life.
Make sure, that the materials are carefully separated.
The electrical conducting boards should be submitted to an appropriate recycling
proceeding.
The rest of the components should be disposed according to the contained
materials.