Mach 380 400 500 700
Dr. Mach
Lamps and Engineering
4.4 Changing of spare parts
4.4.1 Changing the halogen bulbs
Dr. Mach uses special halogen bulbs as illuminants.
Only original Dr. Mach replacement bulbs may be used.
The use of other bulbs can lead to a considerable reduction of the light power and increase in the thermal
load.
The halogen bulbs have a service life of approx. 1000 hours. We recommend changing all halogen bulbs
in a lamp at the same time to keep maintenance work to a minimum.
To replace the halogen bulbs proceed as follows:
•
Turn off the lamp.
•
Loosen the housing cap by pushing the arrest pin
2
(e.g. with a ball-pen).
•
Turn the cover
1
slightly.
•
Remove cover
1
.
ATTENTION: Parts of the housing and the
halogen bulb may be very hot immediately af-
ter use.
•
Loosen the red knurled screw
3
and remove the
lamp socket
4
without removing the electrical
connection
5
.
•
Remove the faulty halogen bulb
6
from the
socket.
ATTENTION: Never touch new halogen bulbs
with your fingers. Always use the original
packaging or a clean cloth.
•
Carefully insert the new halogen bulb
(22.8V/50W).
•
Re-assemble the lamp in reverse order
ATTENTION: When mounting the cover 1, the
arrest pin 2 must snap in audible.
1
2
1
3
4
5
6
59210001
Edition 04