Transferring the colour laser pointer
function to the ring of the handle
In order to transfer the laser pointer function to the
ring of the sterilisable handle, press first push button
12
and then the ON/OFF button on the key pad.
The “LASER” display is blinking.
The previous function on the ring is de-
activated.
A right-turn of the ring at the handle activates the
laser pointer; a left-turn of the ring turns it off.
4.7. Deactivating all functions on the
ring of the handle
The functions on the ring of the handle can be com-
pletely deactivated. Press first of all push button
6
(ENDO)
and then the
ON/OFF button
on the key
pad.
The ring on the handle is without function.
In order to transfer a function on the ring, please
proceed as described in the chapters before.
4.8. Synchronization and / or commu-
nication with the wall panel
The lights Mach LED 2MC/SC can be equipped
against surcharge with the function of synchroni-
zation (combinations only) and / or communica-
tion with a wall panel.
The synchronization of two, three or four OT-
lights enables the transmission of the light inten-
sity or color temperature values set at one light to
the other OT-lights.
Additionally, the communication allows operating
the single lights from a wall panel. The settings
from the wall panel can be transmitted to the
other lights by pressing the „SYNC“ button.
The transmission of the settings to the other
lights is done by pressing the „SYNC“ button on
the key pad.
Step 2
Step 1
Step 1
Step 2
Step 1