TWISTER
®
Softwareversion 1.19
Page 30 of 46
Version 1.19-4 / 2010-05-31
•
The amplitude of the pulse shape is defined as a value between -100 … 100%.
During stimulation it is the percentage of the selected stimulation current in the main
window (cp. Figure 8).
Example:
Amplitude value in pulse shape:
80%
Selected stimulation current in main window:
20mA
Amplitude value in pulse shape [%] * selected stimulation current in main window [mA]
Stimulated current [mA] = --------------------------------------------------------------------------------------------------------------------
100
The stimulated current is 16mA in this case.