K 100 (W) Cl
2
, SCl
2
, ClO
2
, O
2
, O
3
, H
2
O
2
30
.
.
.
Dosage check
S1 10 min
.
Contr. settings
.
.
Basic settings
Main menu
Basic settings
6.8 Dosage check
Controller settings
6.8 Dosage check
In the basic settings of the controller you can define, how long a controller is supposed
to dose with 100% without raising alarm.
If the controller output is 100% for more than the specified time, this is interpreted as
an indication of failure, and the instrument issues an alarm and deactivates the
controller, thus stopping further dosage.
The dosage check is a safety catch to prevent hazardous chemicals to be set free in
case of a defective dosing tube or tube connection.
NOTE
In case of an alarm due to dosage check, only the controller
concerned is deactivated.
NOTE
If you set the dosage check time to 0 seconds, the dosage check
function is deactivated.