Calibration Troubleshooting

If

Err 0

 appears on the display during step 8, the calibration zero is out of range.

Remove all weight from the platform and press LB to perform the zero calibration
again.  If this error persists, consult the raw counts table below.

If

ErrSp

 appears on the display during step 10,12, 14 or 16, the span calibration is

out of range.  Place the proper weight on the platform and press LB to perform the
span calibration again.  If this error persists, consult the raw counts table below.

To Access Raw Counts:
1.  Press and hold ZERO and REWEIGH simultaneously.
2.  When the display reads

Cal 25

 lb (

CAL10

 kg), release ZERO and REWEIGH.

3.  Press KG until the RAW A/D count is displayed.
4.  Apply test weights and compare results to the table below.
5.  Press KG to exit.

The following table shows the acceptable Raw Counts for no load and each
calibration point.  If the scale’s raw counts are not within the ranges specified in the
following table, contact Doran’s Technical Assistance department at
800-264-4107.

Two Point Calibration Troubleshooting

Calibration Requirements in Raw

Counts

Minimum

Maximum

Zero Calibration Point

0

20,000

25 lb (10 kg)

14,000

24,000

45 lb (25 kg)

28,000

42,500

Multiple Point Calibration Troubleshooting

Calibration Requirements in Raw

Counts

Minimum

Maximum

Zero Calibration Point

0

20,000

10 lb (5 kg)

7,000

10,500

20 lb (10 kg)

14,000

21,000

30 lb (15 kg)

21,000

31,500

40 lb (20 kg)

28,000

42,000

Summary of Contents for DS4100

Page 1: ...Model DS4100 Infant Scale Instruction Manual Doran Scales Inc 1315 Paramount Pkwy Batavia IL 60510 USA 1 800 264 4107 www doranscales com MAN0259 Revision 0 5...

Page 2: ......

Page 3: ...eration 5 Section 3 Scale Calibration 6 Two Point Calibration Procedure 6 Multiple Point Calibration 7 Calibration Troubleshooting 8 Section 4 Parameter Setup 9 Entering and Exiting Setup Mode 9 Setup...

Page 4: ...oper handling or abuse can damage the scale and result in costly repairs that may not be covered by the warranty Please observe the following precautions to insure years of trouble free service from y...

Page 5: ...e batteries These batteries supply a year s of typical use with the standard Automatic Shutoff Timer settings The batteries can be accessed on the bottom of the scale To access the battery compartment...

Page 6: ...hutoff Timer that will power down the indicator when not in use Zero When the zero button is pressed the scale will be reset to zero weight This feature can be used to zero out any weight not desired...

Page 7: ...scale status Center of Zero Indicates that the scale has an accurate zero reading Hold Indicates a stable weight has been achieved and held on the display This indicates an accurate weight has been a...

Page 8: ...platform 2 Press On Off The scale will turn on and display a zero weight If the item is removed a negative weight will be displayed Note Each time the scale is powered up the zero point is reset 3 Pl...

Page 9: ...RO and REWEIGH 5 If desired press LB to use a span calibration of Cal 45 lb CAL20 kg NOTE Using a heavier calibration weight will increase scale accuracy 6 Press KG The display reads Cal 0 7 Remove al...

Page 10: ...B button until the display reads CALSP 6 Press KG and the display will read CAL 0 7 Remove all weight from the scale NOTE During calibration be sure vibration and air currents are not present 8 Press...

Page 11: ...AL10 kg release ZERO and REWEIGH 3 Press KG until the RAW A D count is displayed 4 Apply test weights and compare results to the table below 5 Press KG to exit The following table shows the acceptable...

Page 12: ...exit the setup mode press KG until the display returns to the weigh mode Setup Mode Navigation 1 Press KG to change the current parameter 2 Press LB to change the parameter setting Calibration Weight...

Page 13: ...10 seconds 30 30 Seconds 60 1 minute 120 2 minutes 180 5 minutes 300 10 minutes 600 15 minutes of Automatic shutoff timer disabled LB Units Button Enable Disable on LB button active off LB button dis...

Page 14: ...AA Alkaline Batteries One year battery life with 10 second Automatic Shutoff Timer setting Displayed Units lb oz kg Capacity 45 lb x 0 2 oz 20 kg x 5 g Construction ABS Plastic Housing Options Carryin...

Page 15: ...cale Scale will not come to zero when the ZERO button is pressed Make sure that the scale is stabilizing After pressing the ZERO button the scale should zero as soon as it becomes stable Weight readin...

Page 16: ...an calibration is out of range Error is displayed after a span calibration attempt Refer to the Calibration Troubleshooting section for additional information If this problem persists contact Doran Te...

Page 17: ...or indirect damages to equipment or personnel Nor shall Doran Scales Inc be liable for damages to equipment or for personal injury caused by misuse overload accidental damage alteration improper inst...

Page 18: ......

Reviews: