Maintenance
Maintenance UK
It is a legal requirement, under regulation 9 of the COSHH
regulations that all local exhaust ventilation systems are
thoroughly examined and tested at least once every 14
months (typically carried out annually). The approved code
of practice recommends that a visual check should be
carried out at least once a week.
COSHH requires the annual inspection and testing to be
carried out by a competent person and specifies that
documentation results are recorded in a log.
Contact the seller for more information about inspection
and certification.
Maintenance General
User maintenance is limited to cleaning the unit, only the
manufacturers trained maintenance technicians are
authorised to carry out component testing and
replacement. Unauthorised work may result in a potentially
dangerous situation and
will invalidate the manufacturer’s
warranty.
Cleaning the unit
The powder coat finish can be cleaned with a damp cloth
and non-aggressive detergent, do not use an abrasive
cleaning product as this will damage the finish.
The inlets and outlets should be cleaned once a year to
prevent build-up of dust and overheating of the unit.
The Xtract 100 features a removable polycarbonate screen.
This is fitted over the main screen and protects it from any
scratches or abrasions that may occur during the milling
process. The protective screen is disposable and can be
removed by turning the circular latch on the main screen
through 180˚. The screen can now be slid down and
removed from the unit.
Replacement screens are available from BOFA.