Service Note 

 

Before proceeding, discharge any static electricity from your body by touching a bare screw from the 
instrument.  Turn off the instrument and disconnect the AC power cord. 

 
Solution / Action 

Step 1 

Using a Phillips Head Screw Driver, remove the nine screws from underneath the drive plate and 
the one screw located on the top rear of the cover. 

 

 

Step 1 

Step 2

 

Remove the drive plate cover by holding both sides and raising the cover up. 

 

Step 2

 

Step 3

 

Remove the spindle drive belt by pulling it off the idler wheel as shown. 

 

Step 3

 