Service Note
Before proceeding, discharge any static electricity from your body by touching a bare screw from the
instrument. Turn off the instrument and disconnect the AC power cord.
Solution / Action
Step 1
Using a Phillips Head Screw Driver, remove the nine screws from underneath the drive plate and
the one screw located on the top rear of the cover.
Step 1
Step 2
Remove the drive plate cover by holding both sides and raising the cover up.
Step 2
Step 3
Remove the spindle drive belt by pulling it off the idler wheel as shown.