•   Venous  pulsations

•   Fingernail polish or false fingernails 

•   Placement of a sensor on an extremity with a blood 

pressure cuff, arterial catheter or intravascular line

•   High ambient light levels

• 

WARNING

 – The device is intended only as an adjunct in 

resident assessment. It must be used in conjunction with 

other methods of assessing clinical signs and symptoms. 

The device is not intended for continuous monitoring.

• 

WARNING

 – Check skin integrity, circulatory status 

and correct alignment of finger prior to use. Check the 

device sensor application site frequently to determine 

the positioning of the sensor, circulation and skin 

sensitivity of the resident. Do not stretch the lining 

in the opening of the device when inserting finger. 

This may cause inaccurate readings or skin blisters. 

Prolonged use or resident's condition may require 

changing sensor site periodically. 

• 

CAUTION

 – Do not sterilize the device using autoclaving 

or ethylene oxide sterilizing, or immerse the device in liquid, 

which may cause inaccurate readings. The device is not 

intended for sterilization. 

Product Accessories

•  One lanyard

•  One AAA battery

•  One user manual

Directions for Use

Lanyard Installation
1.

  Thread thinner end of the lanyard through the hanging 

hole in the Oximeter.

2. 

 Thread thicker end of the lanyard through the threaded 

thinner end before pulling it tight.

Battery Installation

Install one AAA battery into battery case, following the 

polarity as shown.

Close the battery cover by pushing horizontally along  

the arrow.

• 

WARNING

 – Battery must be installed with the correct 

polarity. Otherwise, DEVICE MAY NOT WORK and damage 

might be caused to device. Remove the battery if the 

device will not be used for an extended period of time.

Operation Instructions
1.

  Insert one finger into the opening of the device. 

When finger is inserted into the device, the fingernail 

surface must be facing upward.

2.

  Press and release the power switch on the front panel  

(If the power switch is pressed and held for more than  

1 second, the brightness of the device will be adjusted.  

There are 10 levels on brightness; the default level is 4.)

3.

  Do not move body or finger while the device is operating.

4.

 Read data from display screen.

Brief Description of Display

Resident pulse rhythm is indicated by the signal intensity 

bar graph.

   

PR

Low 
power

SPO

2

 waveform

   

Six Display Modes

After turning on the device, each time you press and 

release the power switch (for less than 1 second), the 

device will switch to another display mode. There are six 

display modes:

Display Mode

Description

PR

Display mode when turning 

on the device, the display 

is horizontal and upright, 

displaying a solid waveform

PR

 After pressing the power 

button a second time, the 

display is horizontal and 

upright, displaying a line 

waveform

PR

After pressing the power 

button a third time, the display 

is horizontal and upright

PR

SPO

2

PR signal 

intensity Bar
PR

Power
Switch

2