Direct Scientific DSM-502 Operation & Maintenance Manual Download Page 9

 
 
WBJI                                                                                                                                                                        DSM-502 

 

PAGE 9 

 

                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          
6.0 

MAINTENANCE 

 

 

GENERAL:  

The DSM-502 Digital Survey Meter has an Internal Probe that operate independently as  

 

 

 

CPM/CPS RATEMETER, DOSE CALIBRATED RATEMETER, SCALERWITH TIMED  

 

 

 

COUNTING FUNCTIONS .  Once the instrument has been calibrated the only operator  

 

 

 

intervention required is to select the probe with the MAIN  SELECTORS SWITCH ON THE 

 

 

 

FRONT PANEL.   No external devices are required to calibrate the system except a NIST  

 

 

 

traceable radioactive source and electronic pulse generator.  ALL CONTROLS  

 

 

 

 

NECESSARY TO CALIBRATE THE DSM-502 ARE LOCATED ON THE PC BOARDS OF 

 

 

 

THE SYSTEM. 

 

 

6.1   

LOCATION AND FUNCTION OF CALIBRATION CONTROLS   

 
 

GENERAL: 

The calibration procedure for the DSM-502 utilizes only 3 potentiometers and 2 switches to  

 

 

 

calibrate all of the DSM-502 functions. FIGURE 1 SHOWS THE GENERAL LOCATION  

 

 

 

OF THESE CONTROLS.    SW4 located on the left side of the top pc board (next to SW5) 

                 

 

 selects the function to be calibrated.  These switch positions are as follows: 

 
 

POSITION 0    NORMAL OPERATING POSITION 

 

POSITION 1    DISPLAY & EDIT COUNT TIME 0 - 2550 SECONDS (10 SECOND INCREMENTS – ADJ 

–  

 

 

P1 

 

POSITION 2    DISPLAY & EDIT OPERATIONAL ALARM (5% - 95%) ADJ – P1 COARSE – P2 FINE 

 

POSITION 3   DISPLAY & EDIT OVER RANGE ALARM ADJ – P1 COARSE – P2 FINE 

 

POSITION 4    DISPLAY & EDIT CALIBRATION CONSTANT – ADJ – P1 COARSE – P2 FINE – P3  

 

 

 

TENTHS 

 

POSITION 5    DISPLAY & EDIT DEAD TIME - ADJ – P3 

 

POSITION 6    DISPLAY & EDIT HIGH VOLTAGE - MAIN SELECTOR TO #1 PROBE – ADJ #1 HV 

              POSITION 7    DISPLAYS SOFTWARE NUMBER INSTALLED IN INSTRUMENT 
 

POSITION 8    DISPLAY & EDIT FAST & SLOW TC – SET TO FAST ADJ P1 TO (10 – 250 SECONDS)  

                              

SET TO SLOW ADJ P1 TO (10 – 250 SECONDS 

 

POSITION 9    DISPLAY & EDIT LCD REFRESH RATE (8 = 0.8 SECONDS TO 30 = 3.0 SECONDS) 

                              

ADJ P1 TO DESIRED REFRESH RATE  

 
 

SW5                   (located next to SW4) is utilized to ACTIVATE & SAVE CALIBRATION SETTINGS   

                       

 

(

CAL is visible in the upper center of the display when SW5 activates the calibration    

 

 

 

function

) the function selected by SW4 and then save the data once the function has been  

 

 

 

calibrated into the meters data base.   Once a function has been selected by SW4 and  

 

 

 

activated by SW5 the potentiometers P1, P2 & P3 are utilized to input the calibration data.   

 

 

 

Once these potentiometers have calibrated the function SW5 is depressed and the data is  

 

 

 

saved.  This process is repeated for each SW4 position using the same P1, P2, & P3 for all  

 

 

 

data input.   

 
 

 

FIGURE 2 SHOWS THE LOCATION OF ALL THE CALIBRATION COMPONENTS. 

 
 

6.2   

ADJUSTING THE METER FOR CPM/CPS AND R/Sv OPERATION 

 
 

 

6.2.1 

The switches that adjust the DSM-502 for Roentgen or Sieverts are located on the LCD pc  

 

 

 

board. To access these switches remove the DSM-502 from its housing .    Looking at the  

 

 

 

bottom of the instrument with the LCD at the top the CPM/CPS switch is on the left side of  

 

 

 

the LCD pc board. The side of the board near the LCD will have CPS and CPM slide the  

 

 

 

small black switch lever towards the type operation desired. 

  
 

 

6.2.2 

The switch that selects Roentgen or Sieverts is on the right side of the LCD pc board is  

 

 

 

located directly above the push button switch SW5.  Slide the small black switch lever in the 

                              

direction of the desired units of measurement (mR/hr or Sieverts).   

 
 

 

6.2.3 

This completes the adjustment of the DOSE UNITS AND COUNTING UNITS THAT WILL 

 

 

 

BE DISPLAYED ON THE LCD. 

 
 
 

Summary of Contents for DSM-502

Page 1: ...DIGITAL SURVEY METER OPERATIONS MAINTENANCE MANUAL Direct Scientific www drct com tel 310 589 0601 fax 310 919 2981 email sales drct com...

Page 2: ...WBJI DSM 502 PAGE 2 SECTION CONTENTS 1 SPECIFICATIONS 2 DESCRIPTION 3 THEORY OF OPERATION 4 OPERATING CONTROLS 5 OPERATING INSTRUCTIONS 6 MAINTENANCE 7 PARTS LIST 8 SCHEMATICS...

Page 3: ...300 1500 VDC REGULATION 1 LOW VOLTAGE 5 VDC 5 VDC REGULATION 0 5 BATTERY 6 AA ALKALINE BATTERY OPERATION 200 HRS NOMINAL LCD READOUT 4 DIGIT 75 19mm TALL RS 232 DATA OUTPUT 9 PIN MINIATURE SUB D CONNE...

Page 4: ...is switch selectable from the front panel The electronic circuitry is located on 2 heavy duty industrial type printed circuit boards that are directly interconnected to improve reliability and durabi...

Page 5: ...re 3 show the power supply pc board s major components and calibration controls 4 0 OPERATING CONTROLS INDICATORS 4 1 CONTROLS MAIN FUNCTION SWITCH CONTROL SETTING FUNCTION OFF ALL POWER IS DISCONNECT...

Page 6: ...be 5 2 INSTRUMENT OPERATION 5 2 1 Adjust the MAIN SELECTOR switch to RATEMETER for normal ratemeter type operation 5 2 2 Depress the SPEAKER ZERO towards SPEAKER if a audible tick for each input pulse...

Page 7: ...WBJI DSM 502 PAGE 7 FIGURE 2 FIGURE 2 DSM 502 FRONT PANEL CONTROLS P N 12303 REV 6...

Page 8: ...WBJI DSM 502 PAGE 8 FIGURE 3 DSM 502 POWER SUPPLY AND MAJOR COMPONENTS P N 12291 1 REV 1 FIGURE 4 DSM 502 LCD DIGITAL PC BOARD COMPONENTS AND CONTROLS P N 12923...

Page 9: ...IT FAST SLOW TC SET TO FAST ADJ P1 TO 10 250 SECONDS SET TO SLOW ADJ P1 TO 10 250 SECONDS POSITION 9 DISPLAY EDIT LCD REFRESH RATE 8 0 8 SECONDS TO 30 3 0 SECONDS ADJ P1 TO DESIRED REFRESH RATE SW5 lo...

Page 10: ...e MAIN SELECTOR SWITCH at the same setting as was used to disable the OVER RANGE ALARM 6 4 7 Adjust SW4 to position 2 and depress SW5 until the CAL is visible on the display 6 4 8 Adjust P1 P2 until d...

Page 11: ...sed and the CAL is displayed on the readout INTERNAL PROBE DT ADJUSTMENT 6 6 1 Determine the full scale range for the Internal Probe 6 6 2 Be sure MAIN SELECTOR SWITCH is still in RATEMETER position a...

Page 12: ...isappears to save the data Adjust SW4 to the NORMAL OPERATION POSITION 0 ZERO 6 7 5 The survey meter is now ready for NORMAL OPERATION 6 8 ADJUSTING ALARM SETPOINTS GENERAL THE DSM 502 HAS AN AUDIBLE...

Page 13: ...ITCH TO COUNT Adjust SW4 to position 1 This displays the counting time in seconds currently in the system 6 9 2 To set the count time to change COUNT time Depress SW5 until CAL appears 6 9 3 Adjust P3...

Page 14: ...WBJI DSM 502 PAGE 14 6 9 11 If the instrument remains ON a new count cycle can be started by DEPRESSING THE ZERO SWITCH on the front panel for at least 3 seconds...

Page 15: ...WBJI DSM 502 PAGE 15 PARTS LIST...

Page 16: ...100K 5 R2 3 4 5 6 21 R22 23 24 25 R26 27 28 32 14 111143 TERM STRIP 50 PIN J2 10 112729 TRANS 60V 500ma Q1 1 114115 FERRITE CHIP 2 5 A 120 OHM FB2 FB3 2 114206 CRYSTAL 10 MHZ SMT X1 1 114293 SLIDE SW...

Page 17: ...M I C U7 1 109150 RESISTOR 1 6K 5 R6 7 8 9 10 11 6 109156 I C U3 1 109262 RESISTOR 10 MEG 5 R27 1 109263 CAP 22PF 50 V 5 C21 22 25 3 109641 CAP 470 PF 16 V 10 C18 1 109930 52 PIN SOCKET U1 1 110162 FE...

Page 18: ...C24 1 108829 47K 1 8W 1206 PKG R18 1 109011 SOT 23 60V 150A PNP BIPOLAR Q1 1 109133 5V REGULATOR E1 1 109156 MC33172D I C U1 3 2 109163 110K 1 8W 1206 PKG R 6 14 2 109641 470MF 16V C15 1 109896 2 4K 1...

Page 19: ...WBJI DSM 502 PAGE 19 SCHEMATICS...

Page 20: ...WBJI DSM 502 PAGE 20...

Page 21: ...WBJI DSM 502 PAGE 21...

Page 22: ...WBJI DSM 502 PAGE 22...

Reviews: