
D I R E C T H E A L T H C A R E S E R V I C E S . C O . U K
1 4
D Y N A - F O R M A I R P R O - P L U S
Cleaning
Top Cover
Refer to the top cover wash tag for cleaning instructions.
If there are visible signs of body fluids and or substances present, the top cover should be washed. Top covers can be machine
washed (up to 80ºC) using authorised cleaning and disinfection solutions.
To establish the amount of disinfectant to use, determine the amount of water in the washer and then follow the manufacturers’
instructions for dilution.
Soak the top cover in the disinfectant during the wash cycle. Rinse well in clean water and dry thoroughly before use.
Do not dry the top cover using too high a heat cycle (see Dartex technical recommendations - up to 80ºC ). Air dry if possible or
select an appropriate heat dry cycle within limits as above. If there are no visible signs of body fluids and or substances on the top
cover, the top cover should be sanitized and rinsed with fresh water accordingly.
If there are no visible signs of body fluids and or substances on the top cover, the top cover should be sanitized.
1. Apply an intermediate level authorised cleaning and disinfection solution to the top cover upper surface either by spraying or by
hand application.
2. Ensure the surface is completely covered with the disinfectant and remains in contact with the surface according to
manufacturer’s instructions.
3. Remove disinfectant and rinse thoroughly.
4. Allow to air dry before use.
Handle
The exterior of the Handle can be periodically wiped using a cloth and dampened with authorised cleaning and disinfection
solutions.
Control Unit
Ensure the Control Unit is disconnected from the mains electricity supply before cleaning.
Do not spray disinfectant directly on to the Control Unit, or immerse the Control Unit in any type of liquid. This could result in a
severe electrical hazard as this equipment has no protection against ingress of water.
This equipment is not suitable for use in the presence of a flammable anaesthetic mixture with air or with oxygen or nitrous
oxide.
Wipe down Control Unit with warm water containing detergent (or authorised cleaning and disinfection solution) and dry thoroughly
before use.
In case of notifiable diseases clean and disinfect systems following eventually special procedures revised and published by the
local health care authorities. The transport should take place in special plastic bags only.