1. Safety

Dear customer, Thank you for choosing a rehabilitation aid from GATE Rehab. We hope our product meets or 
exceeds your expectations. These Instructions for Use contain the information users of the product need to 
know for its safe operation. In addition to the instructions for use, there is a technical manual that describes the 

chair’s structure and which includes information of a more technical nature and a spare parts list.
 
READ THESE INSTRUCTIONS FOR USE BEFORE USING THE PRODUCT.

This hygiene stool is designed for indoor use on level surfaces and in wet spaces. Max user weight is 150 kg.

The product can be raised, lowered or tilted with the aid of encapsulated, concealed electronics. The product 
is in ingress protection class x5, and can withstand water under a certain pressure. The product may thus NOT 

be immersed or used underwater. This also applies to the frame/castors – may not be immersed in water.

When manoeuvring forward, we recommend use of the directionally lockable castor (right front castor) for 
improved stability. However, the directional lock should not be used when manoeuvring in confined spaces.

The hygiene chair has a fixed battery that is charged via a connection in the remote control. The time required 

for a discharged battery to reach an 80% charge is around 12 hours. The remote control has a charging 
indicator that will show whether or not the battery needs to be charged. Make a habit of always charging the 
battery if at the end of the day it has dropped to level 1 (one green bar on the battery indicator). If the hygiene 

chair will not be used for a few days, we recommend switching off the battery by means of the Off button 

located on the battery.

The hygiene chair also has a service indicator that lights up after around 11 months’ operation as a reminder 
regarding the scheduled service that must be carried out at least once per year. Information about how to reset 
the service indicator can be found in the technical manual.

The hygiene chair has a number of functions to enhance safety. When tilting forward or back, the chair will 

always stop when the horizontal position is reached. The chair cannot be tilted forward if the seat height is 

greater than 65cm. In order to raise the chair higher than 65cm, the seat must be horizontal or tilted to the 

rear. It is not possible to use the chair while charging the battery.

Users will experience a soft start of around 5 seconds when the chair is raised or lowered. This is normal and 

takes place in order to synchronize the two height adjustment actuators.

The hygiene stool is raised and lowered through telescopic pillars with built-in motors. The pillars have an open 

design which allows water and dirt to run through internally. When necessary, clean the pillars and the bottom 

surfaces with cleaning agent (can be purchased from GATE). The cleaning agents dissolve dirt and grease and 
break down any organic substances. Rinse with water. Note that the pillar base plates are accessible for
mechanical cleaning.

Bear in mind the crush risk that is always present when hands and feet get in the way of chair movement, 
especially when tilting forward or when the chair is lowered to its lowest level. Depending on the setting, the 
foot rest may touch the floor or the chair frame.

The staff at GATE Rehab hope you enjoy extensive use of your new hygiene chair. Before you begin using 
your chair, the battery must be activated. Do this by pressing the battery ON button (see section in these 

instructions for use).

Summary of Contents for GATE EASY HighBack

Page 1: ...A DIRECT HEALTHCARE GROUP COMPANY G ATERD S E EASY HighBack User Manual...

Page 2: ...EASY HIGHBACK GATERD SE 2...

Page 3: ...ion concerning EMC 10 EASY in the high back version HB is a development of the conventional concept using the back strap neck support already available This hygiene chair is fitted with a high padded...

Page 4: ...tery by means of the Off button located on the battery The hygiene chair also has a service indicator that lights up after around 11 months operation as a reminder regarding the scheduled service that...

Page 5: ...n and then to the lowest seat height Exercise caution when activating as the system safety sensors are bypassed In particular make sure the chair proceeds evenly to its minimum height to prevent skewi...

Page 6: ...f button located on the battery Make sure the remote control socket charge is always covered plugged in when using the chair Otherwise water may get in and damage the remote control 4 Battery On Of Th...

Page 7: ...features a directional lock see image 15 and the other 3 are fitted with full brakes When the directional lock is activated wheels cannot be turned This is particularly useful when moving heavier pat...

Page 8: ...rd if seat height is greater than 65 cm Before troubleshooting begins always restart the battery see page 4 Hold in the battery s OFF button for approx 5 seconds Wait for one minute Activate the batte...

Page 9: ...an withstand washer water pressure Place the remote control control box or the connector female connector so that running watercannot force its way in Clean the pillars with detergent disinfectants co...

Page 10: ...s functions such that manoeuvring or battery charging fails The hygiene stool includes the following wiring AC cable to the charger 3 wire 2 0 m DC cable between the charger and the remote control 2 p...

Page 11: ...USER MANUAL 11 GATERD SE...

Page 12: ...GATER D SE Gate Rehab Development AB Industrigatan 2 Reftele Sverige S 330 21 T 0371 318 00 F 0371 318 01 E info gaterd se LIT 00080 Issue 1 Date January 2020 A DIRECT HEALTHCARE GROUP COMPANY...

Reviews: