4 • Troubleshooting
Doc. 065207-02 9/08
63
Error 016
Unable to fill cell.
Cause:
An empty (or incorrectly installed) solvent reservoir.
Action:
Refill the solvent reservoir, if necessary. Check that the solvent
reservoir is correctly installed (see
Cause:
A leaking static valve seal.
Action:
Check the collection vessel for liquid. If liquid is present, replace the
static valve seals (see
Cause:
A leaking fitting somewhere in the system.
Action:
Remove the right-side panel of the ASE 150 (see
Inspect the following fittings for leaks (see
): pump
fittings, pump check valves, pressure transducer fittings, pressure
relief valve fittings, static valve fittings, and solvent line fittings. Use
a 1/4-inch open-end wrench (P/N 049452) to tighten any leaking
fittings.
Do not overtighten!
If tightening does not stop a leak, the tubing and fitting assembly
must be replaced. Contact Dionex for assistance.
Cause:
A leaking check valve (internally).
Action:
Replace both check valve cartridges (see
Error 017
Remove cell during oven wait period.
Cause:
A sample cell was installed in the cell holder before the temperature
set point was reached.
Action:
Remove the cell from the cell holder. Never install the cell until the
STATUS
field (on the
STATUS
screen) indicates
OVEN READY
and
the LED on the front panel
START
button is flashing. For details, see
.
Summary of Contents for ASE 150
Page 32: ...ASE 150 Operator s Manual 24 Doc 065207 02 9 08...
Page 64: ...ASE 150 Operator s Manual 56 Doc 065207 02 9 08...
Page 76: ...ASE 150 Operator s Manual 68 Doc 065207 02 9 08...
Page 114: ...ASE 150 Operator s Manual 106 Doc 065207 02 9 08...
Page 132: ...ASE 150 Operator s Manual 124 Doc 065207 02 9 08...