ASE 100 Operator’s Manual
B-12
Doc. 031842-01 5/02
7. When the rinse cycle is complete (after about 1 minute), three beeps
are emitted and the
RINSE
button LED turns off. Remove the rinse
bottle from the system.
If you need to start the next extraction right away, put on the thermal
gloves (P/N 060372) supplied in the ASE 100 Ship Kit (P/N 059397),
remove the cell from the cell holder, and place it on the cell rack
(P/N 059927) to cool.
If you are not starting another extraction right away, open the cell
door and let the cell cool in the cell holder for at least 15 minutes.
Afterward, put on the thermal gloves, remove the cell from the cell
holder, and place it on the cell rack. Close the cell door until the next
extraction, to save energy and prevent anybody from accidentally
touching the inside of the door while it is hot.
Figure B-7. Installing the Rinse Bottle
Summary of Contents for ASE 100
Page 8: ...ASE 100 Accelerated Solvent Extractor vi Doc 031842 01 5 02...
Page 24: ...ASE 100 Operator s Manual 2 10 Doc 031842 01 5 02 Figure 2 4 ASE 100 Rear Panel...
Page 32: ...ASE 100 Operator s Manual 2 18 Doc 031842 01 5 02...
Page 42: ...ASE 100 Operator s Manual 3 10 Doc 031842 01 5 02 Figure 3 2 Inspecting Cell O Rings...
Page 60: ...ASE 100 Operator s Manual 3 28 Doc 031842 01 5 02...
Page 69: ...4 Troubleshooting Doc 031842 01 5 02 4 9 Figure 4 1 ASE 100 Plumbing Diagram...
Page 106: ...ASE 100 Operator s Manual 5 36 Doc 031842 01 5 02...
Page 110: ...ASE 100 Operator s Manual A 4 Doc 031842 01 5 02...
Page 113: ...B Installation Doc 031842 01 5 02 B 3 Figure B 1 ASE 100 Rear Panel...
Page 150: ...ASE 100 Accelerated Solvent Extractor Index 8 Doc 031842 01 5 02...