TPWX2GD/TPWX2GDI/TPW20IEX2GD 4
3. !! WARNINGS !!
For weighing objects made up of light metals (aluminium, magnesium and titanium or their alloys, excluding the alloys
with less than 10% of aluminium, the varnishes and coatings with less than 25% of aluminium in weight), the use of
stainless steel forks is necessary.
The use of the instrument in hazardous areas require a special attention and special precautions during the use and
maintenance
The instrument conforms for use in zones having precise features: do not install and use the instrument in environments
different than those provided for.
The installation, maintenance and repair of the instrument, must be made by qualified and authorized personnel and
with the approval of expert technicians.
The safety of the explosion-proof system is guaranteed
only
if the system is installed, used and taken care of following
the instructions given in this manual and in the technical manual
(TECH.MAN.REF.).
Check that the pallet truck scale is in good operating conditions.
Only spare parts approved by Dini Argeo should be used.
Be very careful when using the instrument; any sparks could cause an explosion.
Avoid accumulations of dust.
To avoid dust layers, the user must regularly clean the surface of the enclosure,
considering that the maximum thickness allowed is < 5mm.
WARNING! ELECTROSTATIC HAZARD, CLEAN ONLY WITH WET CLOTHS OR ANTISTATIC PRODUCT
Avoid accumulations of electrostatic charges; therefore, when using the instrument in a hazardous zone, the appropriate
work clothing must be used by the operator or the maintenance person.
Do not cover the instrument with coverings made by materials which could have electrostatic charge
Do not paint the system
It is forbidden to modify or repair the instrument with components not conforming to the certification; this action
compromises the intrinsic safety of the instrument (
with a subsequent loss of the Ex approval
) and the nullification of
the product warranty.
It is forbidden to connect the instrument to modules not provided for by the certification; this action compromises the
intrinsic safety of the instrument (
with a subsequent loss of the Ex approval
). Contact Dini Argeo srl for further
information.
Avoid direct sunlight.
NEVER load the scale beyond the maximum capacity shown on the weight indicator’s plate.
Avoid the use of oxidising substances (hydrogen peroxide).