Earthing of the system

For correct earthing and optimal system operation, the load cells, junction box, if any, and weighing structure must be earthed.

LOAD CELLS AND JUNCTION BOX

The connection must be made by connecting the earth wires to the earth bar (cables that must have a cross-section of at least 16 mm

2

); 

finally, connect the earth bar to the earth post with a cable having a cross-section of at least 50 mm

2

.

EXAMPLES:

• 

If the load cells are connected to the board through a junction box, the cable shield from the board and the cell cable shields must be 

connected to the earth socket of the junction box (refer to the junction box manual) and the junction box must be earthed using a copper 

cable with a cross-section of not less than 16 mm

2

.

• 

If the load cells are connected directly to the board (without using the junction box), the cell cable shields must be connected to the 

earthing point (or earth bar).

• 

If the weighing system involves large and/or outdoor structures (weighbridges, silos, etc.) and the distance between the junction box 

and the weight board is greater than 10 m, connect the cell cable shields to the earth socket in the junction box. 

WEIGHING STRUCTURE 

Earth the weighing structure and/or any unconnected structures (e.g. silos that release material onto the weighing structure) using cables 

with a cross-section of not less than 16 mm

2

.

Also connect the upper part with the lower part of each cell by means of a copper braid with a cross-section not less than 16 mm

2

 

(refer 

to the 

earthing examples on page 12 and page 13

)

.

SERIAL CABLES AND CONNECTED INSTRUMENTS

Connect the serial cable shield to the earthing point (or earth bar) inside the panel. To avoid any undesired effects, the earth reference 

of the connection cable, power supply and transmitter must be at the same potential.

PC

The 6116EVO is earthed through the power cable.

WARNING: Check that the power supply socket to which the 6116EVO Industrial PC will be connected has an earth line.

GENERAL NOTES:

• 

All earth wires must be of suitable length, so as to obtain an overall resistance of the earthing system of less than 1 Ω.

• 

If the weighing system involves large and/or outdoor structures (weighbridges, silos, etc.): 

- the earth connection must be made by connecting the earth wires to an earth bar and the earth bar to the earth post with 

a cable having a cross-section of not less than 50 mm

2

- the thickness of the cables must be greater (50 mm

2

 instead of 16 mm

2

 and 100 mm

2

 instead of 50 mm

2

), because the 

voltages at stake are greater (e.g. lightning); 

- the earth post must be placed at a distance of at least 10 m from the structure.

• 

If the load receiver is more than 10 m from the transmitter, we recommend using the SENSE line and load cells equipped 

with a (SENSE) compensation circuit.

Summary of Contents for 6116EVO

Page 1: ...www diniargeo com 6116EVO Industrial touchscreen PC with integrated scale input ENGLISH USER MANUAL...

Page 2: ......

Page 3: ...lation 5 Installation requirements 5 Electrical precautions 5 Earthing of the system 7 Dimensional technical drawings 8 6116EVO 6116EVO PC 8 6116EVO DGT 9 Wiring diagrams 10 6116EVO 10 6116EVO DGT 11...

Page 4: ...y the greatest attention to all those paragraphs that illustrate the simplest and safest way to operate The utmost care has been taken in compiling this manual but reports of any inaccuracies are alwa...

Page 5: ...d in this manual constitutes improper use of the equipment Electrical precautions Use a regulated mains supply within 10 of the rated voltage The electrical protections fuses etc are the responsibilit...

Page 6: ...tres with 2 x 0 5 mm2 cable 150 metres with 2 x 1 mm2 cable RECOMMENDED DISTANCES AND CABLE CLASSIFICATION MAXIMUM ALLOWED LENGTHS Category I Category II Category III Category IV Distance 100 mm 200 m...

Page 7: ...t release material onto the weighing structure using cables with a cross section of not less than 16 mm2 Also connect the upper part with the lower part of each cell by means of a copper braid with a...

Page 8: ...8 Optimized layout for A4 print 329 100 100 66 233 211 327 379 USB ETHERNET VESA 100 M4 x 10 mm PG9 2 7 26 140 7 11 151 354 6116EVO 6116EVO PC USER_MAN_ENG_6116EVO Dimensional technical drawings...

Page 9: ...9 Optimized layout for A4 print 329 100 100 120 120 66 15 5 233 211 327 379 USB ETHERNET PG9 151 354 2 7 26 140 7 11 6116EVO DGT USER_MAN_ENG_6116EVO...

Page 10: ...timized layout for A4 print RS485 Cell 4 Cell 3 Cell 2 Cell 1 Power 485 POWER CELL ETHERNET USB 2 0 USB 2 0 POWER CELL 3 CELL 2 ETHERNET CELL 1 USB 2 0 USB 2 0 6116EVO USER_MAN_ENG_6116EVO Wiring diag...

Page 11: ...11 Optimized layout for A4 print POWER CELL ETHERNET USB 2 0 USB 2 0 6116EVO DGT USER_MAN_ENG_6116EVO...

Page 12: ...SUPPLY CELL 1 CELL 4 CELL 3 CELL 2 B Vdc GND Vdc GND EXC EXC Sen Sen Sig1 Sig1 Exc Exc Sig2 Sig2 Exc Exc Sig3 Sig3 Exc Exc Sig4 Sig4 K1 K2 LOAD CELL 1 E E S S E E S S SIG SIG SEN SEN EXC EXC LOAD CELL...

Page 13: ...AL 4 Sig4 LOAD CELL E E S S SIG SIG SEN SEN EXC EXC LOAD CELL 2 4 SIG 12 24 Vdc 18 17 16 6 7 8 3 4 5 1 2 13 12 15 14 11 10 LOAD CELL RS232 RS485 POWER B A 9 GND RX TX DISPLAY J1 CAL J1 M DGT1 USER_MAN...

Page 14: ...ntrol WORKING TEMPERATURE 0 60 C 0 40 C for type approved applications 6116EVO DGT only DIMENSIONS Refer to the dimensional technical drawings POWER AND CONSUMPTION 110 240 Vac max 75 W OIML SCALE INP...

Page 15: ...Sig1 Sig1 Exc Exc K1 SW1 CELL 3 Exc Sig4 Sig4 LD1 SW1 1 CELL 3 CELL 3 CELL 2 Sig1 Sig1 Exc Exc Sig2 Sig2 Exc Exc Sig3 Sig3 Exc Exc Sig4 Sig4 A B A TERM RS485 PWR SUPPLY CELL 1 CELL 2 B Vdc GND Vdc GND...

Page 16: ...FIELDBUS 25 GND 16 15 14 13 12 11 CELL SIG EXC EXC SIG SIG SEN SEN 10 USER_MAN_ENG_6116EVO 6116EVO DGT The transmitter has a channel CELL for connection to 6 wire load cells with the use of SENSE The...

Page 17: ...le takes place through ASCII serial protocol or Modbus RTU Check the serial port to be used for communication using the Windows Device Manager program in the Ports COM LPT submenu For more information...

Page 18: ...ves the right to make changes to its products at any time without notice and without incurring any penalty We therefore recommend that you always check for updates The person responsible for the use o...

Page 19: ......

Page 20: ...536 921784 Fax 39 0536 926654 www diniargeo com The information in this document is approximate and can be subject to variations without prior notice by Dini Argeo in compliance with the norms in forc...

Reviews: