Ultrasonic level meter ULM-55 - instructions

I. Description

 

The ULM-55 ultrasonic level meter is designed for continuous non-contact measurement of the level of liquids 

and mushy and paste-like substances in open and closed vessels, reservoirs, sumps etc. Depending on the working 
area, the N type can be applied in environments not exposed to the danger of explosion, whereas the Xi type is 
intended for use in explosive environments.
Table 1 contains classifi cation of the environments.
The adjusting elements of the level meter are situated on the rear side of the sensor, under the cap nut (see 
chapter V).

Table 1: Meter type and description of working areas (according to the Norms EN 60079-14, EN 60079-10)

ULM-55N           

                                                                        

                  non-explosive area

ULM-55Xi         II 1/2G Ex ia IIBT5          with isolating repeater (e.g. IRU-420) 

complete level meter     zone 1

  

                 

                                                                        

front part of head          zone 0

 

The level meter ULM-55 transmits by electro-acoustic transducer the series of ultrasonic pulses, which propa-

gate towards to the level. Refl ected acoustic waves are received back by the transducer and processed in the 
electronic  module. The echos are traced in the fl oating time window, the time of the wawes fl ight are fi ltered and 
averaged. Further is made the thermal compensation and the signalis transferred to the output current 4÷20 mA. If 
continuing echo fault happens, the current of the level meter slowly falls down to the value <3,8 mA. When the level 
is moving in the dead zone (see the catalogue’s  data) the output current rises up to the alarm value >21 mA. This 
value is kept, untill the level meter doesn´t register the continuous motion of the level away from the dead zone to 
the nominal measuring range (jumping distance is evaluated as a permanent fault).

II.  Assembly and disassembly

 

Level meters are installed in vertical position into the upper lid of a reservoir or a vessel, using welded-on piece, 

nut or fl ange. The axis of the level meter must be perpendicular to the surface of the liquid being measured. The 
installation point must be chosen in such way so that the emitted acoustic signal is not affected by near objects (re-
inforcements, ladders, mixing devices etc), liquid being poured inside, air fl ow etc. If in doubt, we advise to consult 
it with the producer. The level meter housing need not to be grounded. Recommended dimensional parameters for 
the level meter installation are shown in fi gure 1.

III. Installation

 

Terminals for electric connection are inside the dismountable GDM connector.  Connect the positive pole (+) 

Figure 1: Recommended dimensional parameters - dimension "a" must be as minimal as possible

b > 100 mm
a / b < 1,5
c > 2000 mm
c / d < 8

b > 100 mm
a / b < 3
c > 800 mm
c / d < 8

b > 100 mm
a / b < 3
c > 500 mm
c / d < 8

ULM-55_-02

ULM-55N-10

ULM-55_-06

Summary of Contents for ULM-55N

Page 1: ...nalis transferred to the output current 4 20 mA If continuing echo fault happens the current of the level meter slowly falls down to the value 3 8 mA When the level is moving in the dead zone see the catalogue s data the output current rises up to the alarm value 21 mA This value is kept untill the level meter doesn t register the continuous motion of the level away from the dead zone to the nomin...

Page 2: ...dure 1 Level meter is connected to the supply line via a miliampere meter controller etc 2 Disconnect the connector for access to adjusting elements and unscrew the cap nut mind the internal wires Connect again the connector 3 Empty the tank to the minimum level or put the level meter to the corresponding distance Set the static current by 4 mA trimer to 4 mA Turning to the right clockwise increas...

Page 3: ... 4 mA VI Usage operation and maintenance The level meter requires no operating personnel In the case of operation the optional local connected display informs operating personnel about the measured level The maintenance of the unit includes a check up on the good condition of the level meter and the supply cable Depending on the nature of the measured substance we recommend to inspect the cleanlin...

Page 4: ...rsion Year of production first two digits and serial number remaining three digits xx xxx Ex labelling in the hexagon rgerhgType II1 2GExiaIIBT5 Number of the spark safety certificate Limit operating parameters Ui 30 VDC Ii 132 mA Pi 0 99 W Ci 370 nF Li 0 9 mH Range of head working temperatures ta 30 70 C ta 20 60 C zone 0 Protection IP65 IP67 CE labelling Number of the authorised person inspectin...