5.4 States & Messaging of NIR analyzer internal system
After turning on the X-
NIR™, before going on with the execution of the analysis it is better to
wait always at least
15 minute
before starting to allow :
the necessary warming up of the internal lamp;
one first calibration of the internal optical of the NIR analysis system.
To assist the user the X-
NIR™ system always displays at the top right of its screen an icon
which shows the current status of the X-
NIR™ NIR analysis system:
X-
NIR™
Description of the icon
WARM-UP
the NIR system is executing the initial warm-up phase of the internal
lamp.
OPTICAL CALIBRATION
the NIR system is executing the calibration phase of the internal optical.
READY
the X-
NIR™ system is ready for the analysis execution.
RUNNING ANALYSIS
the X-
NIR™ system is executing the analysis.
CONNECTION MISSING
or
ERROR STATUS
the NIR group is not connected to the electronic unit or there is an issue
the internal NIR system.
ERROR OF NIR CALIBRATION
The system cannot perform the calibration.
After turning ON the machine
BEFORE STARTING
with the analysis executions, the X-
NIR™
system executes automatically :
1.
the
WARM-UP phase
of the NIR group with status icon
, to allow the necessary
warming of the lamp inside the NIR Reader;
2.
the
FIRST CALIBRATION phase
of the NIR group with status icon
, to allow the first
calibration of the internal optical of the NIR analysis system;
3.
after these first two phases the X-
NIR™ analyzer is
READY
, with status icon
.
Although the WARM-UP takes less time and the CALIBRATION phase is
already done, i
t’s STRONGLY RECOMMENDED to wait always at least
15
minutes
before starting use the X-
NIR™ machine in order to allow the
internal lamp to warm up.