17
40.0
℃
/104.0
℉
for at least 30 minutes
before taking a temperature
measurement.
3.If the problem persists after trying
the above method, please to contact
the dealer or manufacturer.
Err
Environment
temperature is not in
10
℃
to 40
℃
or 50
℉
to 104
℉
range.
1.Firstly make sure the probe is clean
and no dirty on it.
2.If it`s the enviroment tempertature
out of range, then patient and
thermometer should both stay in a
room between 10.0
℃
/50.0
℉
and
40.0
℃
/104.0
℉
for at least 30 minutes
before taking a temperature
measurement.
3.If the problem persists after trying
the above method, please to contact
the dealer or manufacturer.
9.
Specification
Product name
Medical infrared Thermometer
Product model
PC809/PC809EV/PC809V
Applicable
regulations and laws
ASTM E 1965 / EN12470-5 / GB/T 19146
Power supply
DC 3.0V (2 pcs of AAA battery)
Battery life
More than 2000 measurements
Low battery indicator
Appears on screen when the voltage is below 2.6V
±
0.1V
Measurement mode
Forehead& Ear
Measurement time
1s
Temperature units
℃
&
℉
, convertible
Measurement range
32.0
℃/89.6℉ to 42.9℃/109.2℉
Accuracy
±0.2
℃/0.4℉
Display resolution
0.1
℃/℉
Automatic shutdown
30s
±
1s
Memory
32 sets of measurement
Operation environment
Temperature: 10.0
℃
/50.0
℉
~ 40.0
℃
/104.0
℉
Humidity: 20%-85% RH, non-condensing
Atmospheric pressure:86kPa ~ 106kPa
Storage & shipping
environment
Temperature: -20
℃
/-4
℉
~ 55
℃
/131
℉
Humidity: 20%-93% RH, non-condensing
Atmospheric pressure:86kPa ~ 106kPa
Date of manufacture
See the label