background image

Model:

  

 

2024T 

 
Sensor: 

 

  

 

PT100 Class A 

Connector Type:

  

 

4 pin TCM screw plug  

Units: 

 

  

 

°C/°F 

Operating Temperature:

            -10°C to +50°C 

Operating Range:

 

             -196°C to +600°C 

Accuracy of Unit Only

:    

± 1 Digit 

( 0.1˚C Low scale; 1˚C High scale )

 

 

 

 

±0.2% of reading 

±0.1˚C over range of 

-

196˚C to +600˚C

  

 

Before Use 

Please ensure this unit is suitable to application.  
Check temperature range and thermocouple type to ensure suitability for application. 
 

Power 

The instrument requires two AA or equivalent cells. These are not supplied with the unit. 
When the batteries require changing the following symbol will be displayed:  
Please ensure the screws holding the battery compartment are firmly tightened when inserting or replacing batteries 
to ensure the IP65 rating is maintained. 
Press ON key to turn unit ON. Press OFF key to turn unit OFF. 
The unit will switch OFF automatically after 12 minutes, unless any key is activated. 
To disable automatic switch off, hold      Key when the unit is being switched on. This will disable the automatic 
switch-off function until the unit is switched off. 

 
Hold Function 

To enable the Hold Function, press the “Hold”

        button  

 

Speedread 

This is a function that gives an indicative reading within 14 seconds for poor thermal conductivity items. The use of 
this function 

will reduce the accuracy to ±2˚C. For correct Speedread measurements the probe must be in good 

thermal contact with medium under test prior to activating the function. A steep temperature gradient or wide 
temperature and item under test are required. Speedread is suitable for core frozen food measurements and 
hot/cold liquids. Good thermal contact is unlikely when a general purpose or between pack probe is used on the 
surface of food produce. (Not suitable for use with between pack probe). 
Insert probe and press   key. The   symbol will appear and the display will flash fast until the correct reading 
appears and the     symbol will be displayed. This reading will automatically be held. Press     key to return to normal 
mode. If it is not possible to obtain a faster reading due to the similarity of probe tip temperature and the 
temperature of item to be measured, then the   symbol will be cancelled and the instrument will return to normal 
mode. 
 

C/F˚C Selection

 

Select the desired temperature scale by pressing 

key. The selected scale is confirmed by ˚C or ˚F symbol on the 

display, and is retained when the instrument is switch off.  
 

Range lock 

All instruments have a 0.1˚ resolution from 

-

199.9˚C to+199.9˚C and a 1˚ resolution outside this range. To fix the 

di

splay to a 1˚ resolution, press the 

A/R

 key. To cancel the mode, press the 

A/R

 key again and A/R will be shown on 

the display. 

 

 
 
 
 
 
 