DIGISYSTEM LABORATORY INSTRUMENTS INC.
Service Manual
Troubleshooting Guide
Caution:
Always disconnect the power cord before troubleshooting.
Trouble / error message
Cause
Remedy
Motor doesn
’
t work
(Error message :
Er-P
)
1.
Speed sensor is defective.
2.
Motor is defective.
3.
Motor driver PC board is defective.
4.
Carbon brush is worn out.
1.
Contract the distributor for repair.
2.
Contract the distributor for repair.
3.
Contract the distributor for repair.
4.
Contract the distributor for repair.
Motor is overshooting.
(Error message:
Er-E
)
1.
Speed sensor is defective.
2.
Motor is defective.
3.
Motor driver PC board is defective.
1.
Contract the distributor for repair.
2.
Contract the distributor for repair.
3.
Contract the distributor for repair.
Lid open error.
(Error message :
Er-d
)
1.
During operation, the lid is open.
1.
Waiting for the motor stop
completely. And close the lid.
Speed value is not correct.
(Error message:
Er-r
)
1.
Speed sensor is defective.
2.
Motor is defective.
3.
Motor driver PC board is defective.
1.
Contract the distributor for repair.
2.
Contract the distributor for repair.
3.
Contract the distributor for repair.
Error message:
run
1.
When the motor is rotating, push
the button .
1.
Waiting for motor stop completely,
and then push the button ,
and then close the lid to restart the
centrifuge.
No display on the screen
1.
Power is off.
2.
Dead power output
。
3.
Burnt fuse
4.
PC board is defective.
1.
Turn on the power.
2.
Change to a different output.
3.
Replace a new fuse.
4.
Contract the distributor for repair.
Burnt fuse
1.
Fuse is wasted
2.
Motor is defective.
3.
PC board is defective.
1.
Replace a new fuse.
2.
Contract the distributor for repair.
3.
Contract the distributor for repair.
Lid can
’
t be open
1.
Lid lock is defective.
2.
PC board is defective.
3.
PC board is crashed.
1.
Contract the distributor for repair.
2.
Contract the distributor for repair.
3.
Turn off the power, and turn on the
power again. Push the button to
open the lid.
The rotor is slanted
1.
Motor mount is worn-out.
1.
Contact your distributor for repair.