DIGISYSTEM LABORATORY INSTRUMENTS INC.
Circuit Diagram:
Troubleshooting Guide
Caution:
Always disconnect the power cord before troubleshooting.
Trouble
Cause
Remedy
Instrument inoperative
Power cord not connected to outlet.
Plug instrument in..
Dead power output.
Change to different output.
Current is overload, so the no fuse
breaker on the behind of the controller
Push the button of the no fuse
breaker to reset
Control head can’t control
the water temperature
Sensor is broken
Replacing a sensor
PID controller
Replacing a PID controller
Heater is broken
Replacing a heater
Pump Circulator doesn’t
work well
Pump Circulator is broken
Replacing a Pump Circulator
Foreign bodies blockaged the pump
Circulator
Sweep away the foreign bodies
The key of PID controller
can’t work
The keys has been locked
Please refer to page 2, Other
functions 5. (c) to release the
lock status.
Temperature is not stable The initial setting value is not suitable
for the ambient air temperature where
the machine is located
Please refer to page 2, Other
functions 4. to set auto-tuning
function to solve this problem.
PV value is not correct
User’s calibration thermometer is
different from the factory’s calibration
thermometer
Please refer to page 2, Other
functions 3 to adjust PV value