Circuit Diagram: 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Troubleshooting Guide

 

Caution:

 

Always disconnect the power cord before troubleshooting.

 

Trouble 

Cause 

Remedy 

Instrument inoperative 

Power cord not connected to outlet.   

Plug instrument in.. 

Dead power output. 

Change to different output. 

Current is overload, so the no fuse 
breaker on the behind of the controller 

Push the button of the no fuse 
breaker to reset 

Control head can’t control 
the water temperature 

Sensor is broken 

Replacing a sensor 

PID controller 

Replacing a PID controller 

Heater is broken 

Replacing a heater 

Pump Circulator doesn’t 
work well 

Pump Circulator is broken 

Replacing a Pump Circulator 

Foreign bodies blockaged the pump 
Circulator 

Sweep away the foreign bodies 

The key of PID controller 
can’t work 

The keys has been locked 

Please refer to page 2, Other 
functions 5. (c) to release the 
lock status. 

Temperature is not stable  The initial setting value is not suitable 

for the ambient air temperature where 
the machine is located 

Please refer to page 2, Other 
functions 4. to set auto-tuning 
function to solve this problem. 

PV value is not correct 

User’s calibration thermometer is 
different from the factory’s calibration 
thermometer 

Please refer to page 2, Other 
functions 3 to adjust PV value 

 
 