Operating instructions

3.1

Mains input

The mains input connector is located on the rear of the unit. Before connecting to the mains supply the user
should verfy that the unit is set to the correct voltage for the country of use. The mains voltage is set by the
external mains voltage selector switch located on the rear of the unit.

3.2

Mains input fuse

The mains fuse is located on the rear of the unit. The correct fuse for each supply voltage is as follows :

Model

110Vac

220Vac

HY3003

T4A 250V

T2A 250V

HY3003-2

T6.3A 250V

T3.15A 250V

HY3003-3

T6.3A 250V

T3.15A 250V

HY3010

T10A 250V

T5A 250V

HY3020

T15A 250V

T8A 250V

3.3

Power on/off

The mains power to the unit is switched on and off using the POWER on/off pushbutton. 

3.4

Voltage and current display

Single output models - HY3003, HY3010 and HY3020 feature independent displays for both output voltage
and output  current. Multiple output  models - HY3003-2 and HY3003-3 feature a single display for each
adjustable output and can be switched to display either output voltage or output current.

3.5

Single output unit or multple output unit in Independent mode

For multple output units, Independent mode is set by selecting the two Tracking switches to the Independent
postion. The output voltage is adjusted to the desired value using both the Coarse and Fine (if fitted) Voltage
controls. The output current level is adjusted to the desired value usng both the Coarse and Fine (if fitted)
Current controls. If the output current exceeds the set current value the power supply will limit the output
current to the set value.

3.6 

Multiple output unit in Series mode

Multple output units can be set to Series mode by selecting the two Tracking switches to the Series position.
The output voltage is set by the Master Voltage control and the Slave output will track the Master output
voltage level. Current adjustment is independent and requires the Slave output current setting to be high
enough otherwise tracking will not occur as the Slave output will current limit. In this mode the +ve terminal
of the Slave output should be connected with a thick wire to the –ve output terminal of the Master to avoid
damage to the unit in the event of an overload. In this mode the maximum output voltage is the sum of the
Master and Slave output voltages.

6

Summary of Contents for HUC 42-01

Page 1: ...1 HY3003 HUC 42 01 HY3003 2 HUC 42 02 HY3003 3 HUC 42 03 HY3010 HUC 42 04 HY3020 HUC 42 05 digimess GEBRAUCHSANWEISUNG OPERATING INSTRUCTIONS Bestell Nr Order No NETZEIL POWER SUPPLY HY SERIES...

Page 2: ...Nachdruck auch auszugsweise nur mit schriftlicher Genehmigung des Herstellers Alle Rechte vorbehalten Subject to alterations errors excepted Reprints also in extracts are only allowed with written per...

Page 3: ...vered and a free space of about 10 cm should be ensured 1 2Switching on The power supply is switched on using the power switch at the front The power switch separates the unit completely from the prim...

Page 4: ...we will of course be glad to repair your power supply as per our General Terms Of Assembly And Service 1 6Description The digimess HY series is a comprehensive range of low cost general purpose linear...

Page 5: ...regulation 0 01 5mV 0 01 5mV 0 01 5mV 0 02 10mV 0 02 25mV Ripple noise 1mVrms 1mVrms 1mVrms 3mVrms 3mVrms Display 2 x 3 5 digit LED Accuracy V 1 0 2 dig I 1 5 2 dig 2 x 3 5 digit LED Accuracy V 1 0 2...

Page 6: ...t 3 5 Single output unit or multple output unit in Independent mode For multple output units Independent mode is set by selecting the two Tracking switches to the Independent postion The output voltag...

Page 7: ...In this mode the ve terminal of the Master output should be connected to the ve terminal of the Slave output the ve terminal of the Master output should be connected to the ve terminal of the Slave ou...

Page 8: ...kt sein und ein Spielraum von ca 10 cm sollte gew hrleistet sein 1 2Einschalten Das Netzteil wird mit dem Stromschalter vorne eingeschaltet Der Stromschalter trennt die Einheit v llig von der Prim rse...

Page 9: ...dlich gerne bereit Ihr Netzteil gem unserer All gemeinen Montage und Servicebedingungen zu reparieren Beschreibung Die digimess Einheiten der Serie HY umfassen ein umfangreiches Angebot von kosteng ns...

Page 10: ...1 5mV 0 01 5mV 0 01 5mV 0 02 10mV 0 02 25mV Brumm St rung 1mVrms 1mVrms 1mVrms 3mVrms 3mVrms Display 2 x 3 5 stell LED Genauigkeit V 1 0 2 stell I 1 5 2 stell 2 x 3 5 stell LED Genauigkeit V 1 0 2 ste...

Page 11: ...it mit mehreren Ausgaben im sog Independent Modus unabh ngigen Modus Bei Einheiten mit mehreren Ausgaben wir der Independent Modus durch Stellen der zwei Nachlaufschalter Tracking auf die Position Ind...

Page 12: ...e Nebenstromregelung hat keinerlei Auswirkung In diesem Modus sollte der positive Anschluss der Hauptausgabe an den positiven Anschluss Nebenausgangs angeschlossen werden der negative Anschluss des Ha...

Page 13: ...H o u s e S t e n s o n D e r b y D E 7 3 1 H L E N G L A N D N e t z e i l P o w e r s u p p l y B l o c d a l i m e n t a t i o n H Y 3 0 0 3 H Y 3 0 0 3 2 H Y 3 0 0 3 3 H Y 3 0 1 0 H Y 3 0 2 0 H U...

Reviews: