Replaceable Sensor Certificate of Validation

Replaceable Sensor Certificate of Validation

Replaceable Sensor Certificate of Validation

Replaceable Sensor Certificate of Validation    

 
 

Dickson ensures that the Replaceable Sensors / Instruments listed below were developed, tested 
and validated together. 
 

Replaceable Sensor Model: R200 and R250 

 

Instrument Model: TH8P0, TH8P2, TH8P3, TH8P5  

 
 
Operation

Operation

Operation

Operation:  The Instruments and Replaceable Sensor models listed above were specifically 
designed to work together.  No other sensors will work with the Instrument models listed.  No 
other Instrument will work with the Replaceable Sensors listed.  The Replaceable Sensor must be 
connected to the Instrument for the Instrument to operate properly. 
 
Calibration

Calibration

Calibration

Calibration:  Only the Replaceable Sensor is calibrated.  The sensor, and all calibration defaults 
and adjustments are stored on the Replaceable Sensor.  Accuracy is strictly controlled by the 
sensor – no adjustments are made by the Instrument.  Readings are sent from the Replaceable 
Sensor to the Instrument for storage purposes only.  Certificates of Calibration / NISTs are only 
supplied for the Replaceable Sensor for this reason.   
 
Recalibration

Recalibration

Recalibration

Recalibration:  When a Replaceable Sensor is due for recalibration the existing Replaceable Sensor 
can be replaced with a new Replaceable Sensor thus eliminating the need to return the entire 
instrument for recalibration.  Replaceable Sensors can be returned for recalibration as well. 

Summary of Contents for TH8P0

Page 1: ...a Set the chart time manually by inserting a coin into the groove in the chart hub and turning clockwise until the correct hour and day if applicable on the chart is referenced to the timing arrow jus...

Page 2: ...ed on the outside edge of the chart press the Pen Home Pen Home Pen Home Pen Home to move pen s to recording position If the pens are located on the chart press the Pen Home Pen Home Pen Home Pen Home...

Page 3: ...n will scroll through the five options until the A A A Alarm larm larm larm button is pressed to exit from alarm adjust Each press of the Pen Home Pen Home Pen Home Pen Home or Alarm Alarm Alarm Alarm...

Page 4: ...Use a Pen or small screw driver to flip the switches Remember to install the correct chart to match the corresponding Dip Switch setting Re Re Re Recording Time cording Time cording Time cording Time...

Page 5: ...s over time 1 While the unit is on press and hold both Pen Home Pen Home Pen Home Pen Home and On Off On Off On Off On Off buttons until the LED is solid green Release the Pen Pen Pen Pen Home Home Ho...

Page 6: ...th the On Off On Off On Off On Off and the Adjust Adjust Adjust Adjust Down Down Down Down button until the LED is solid green The LED will then blink amber at which point only the pen being adjusted...

Page 7: ...al Why does the display read PROB Why does the display read PROB Why does the display read PROB Why does the display read PROB Check the TH sensor pod Make sure it is properly plugged into the unit Is...

Page 8: ...won t the unit power up Remove the batteries and power adapter for a minute or two this will reset the unit The unit should respond when the adapter is plugged in WARRANTY WARRANTY WARRANTY WARRANTY D...

Page 9: ...The Replaceable Sensor must be connected to the Instrument for the Instrument to operate properly Calibration Calibration Calibration Calibration Only the Replaceable Sensor is calibrated The sensor...

Reviews: