  Source or Transmitter

wire of the transmitter to the negative terminal of the 

appropriate channel.  Tighten screws in terminal block 

to hold in transmitter wires.  Repeat steps with second 

transmitter.

  Any powered transmitter or battery source with a 

0-15Vdc output signal may be used.  To connect a 

transmitter to the logger, locate the terminal block on 

the side of the logger.  Connect the positive wire of 

the transmitter to the positive terminal and the negative 

  Battery Replacement

contact the factory if the battery requires replacement.  

The Battery Level Monitor on the Setup screen displays 

the battery voltage and a Low Battery Warning when 

replacement is needed.

  The battery will operate the logger for approximately 

5 years at room temperatures.  Operation of the logger 

at the temperature extremes, in Real-time mode for 

prolonged periods, or sample intervals of less than a 

minute, will significantly reduce battery life.  Please 

  Factory Service and Returns

NOTE:

 Dickson shall not be liable for consequential or 

incidental damages resulting from failure or malfunction 

of its products.

  Contact the factory for a Return Authorization (RA) 

Number before returning any instrument.  The model 

number, serial number and a purchase order number 

will be requested before an RA number is issued.

+

-

+

-

+

-

Power Supply

Transmitter 1

Transmitter 2

ES120-A

CH1

CH1
CH2
CH2

IN Voltage

SIG OUT

GND

IN Voltage

SIG OUT

GND

Typical Setup,

Voltage (V)

  Indicator & Troubleshooting

  • Verify transmitter is functioning properly.  Try 

    another transmitter.  Verify wiring polarity and 

    wire connection is secure.  These loggers 

    are designed for use in a non-condensing 

    environment.  If a logger is going to be used 

    in a condensing environment, try placing the 

    unit in a small sealed plastic bag to protect it from 

    condensation.

  • If the logger appears to be malfunctioning, 

    try the following: open the unit by removing the 

    screw in the back of the case.  Press the small 

    Reset button located on the circuit board 

    inside.  Try the logger operation again.  

NOTE

    Pressing the Reset button will stop data logging.  

    Download the data and Clear memory to start 

    logging again.

  • 

There is a red indicator located on the front 

    lower right of the unit.  This indicator will 

    flash every 10 seconds when the unit is sampling 

    and operating properly.  If the indicator is not 

    flashing, please verify the following:

    • The logger memory may be full; download the 

      data and Clear Memory.

    • Sampling may not be started; download the data 

      and Clear Memory.

    • Condensation may have formed on unit.  Let it sit 

      in a warm dry environment for 24 hrs.  Download 

      the data, clear memory and try again.

  • The battery may require replacement; download 

    the data and check the Battery Level Monitor on 

    the Samples tab of the Setup screen.  If unable to 

    communicate, the battery voltage may already be 

    low.