REMOVING THE USED LANCET
CLEANING AND DISINFECTION OF THE PRICKER
PRICKING IN THE ALTERNATIVE SITE
1.
Remove the pricker adapter
Remove carefully the adapter from the
pricker.
Be careful not to touch the
uncovered lancet.
1.
Place the lancet
Place the lancet in lancet carrier; press it to
the end inside the appliance.
2.
Install the transparent adapter for the
alternative site testing (AST).
Install the transparent adapter for the
alternative site testing (AST) in place of the
pricker adapter.
The adjustment of the puncture depth is not
required if the (AST) adapter is used.
3.
Pull the sleeve
Pull the slide-on sleeve forward until the
click is heard.
4.
Hold down the transparent adapter on
the skin
Press down the pricker stably and then
press the release button until the sufficient
blood volume appears in the adapter.
5.
Lift up carefully the pricker vertically.
Be careful not to spread the blood sample at
the puncture site. Apply a drop of blood on
the test strip.
2.
Put the cover on the lancet
Before the removal of the lancet, put the
protective lancet cover on the hard surface,
then stab carefully the lancet in it.
3.
Remove the lancet
Slide the lancet ejector forward until the
lancet pulls out of the pricker. Remove the
used lancet.
4.
Install the adapter
Reinstall the pricker adapter on the
appliance.
IMPORTANT:
If you consider the use of the other puncture
site than the fingertip,
consult a physician.
The transparent adapter for pricking in the alternative sites
(forearm or palm of the hand) is designed for the use with
Optilet Mini in order to collect a blood sample for measuring
the concentration of glucose.
For use only by a single patient.
Read the instructions for use of glucose meter in relation to the
collection of blood samples from the alternative sites.
In order to clean the pricker, wipe the elements with a soft cloth wet
with water and mild detergent. Do not submerge the appliance in any
liquid. In order to disinfect the elements, prepare a solution of
household bleach and water in the ratio 1:9. Wipe the appliance with
a soft cloth wet with this solution. Submerge only the adapter in this
solution for 30 minutes. After disinfection, rinse and let it dry in the
air.
FUNCTION OF PRECISE CLEANING
If blood enters the interior of the pricker adapter, turn the
transparent pricker tip to the position "
↑
" (Fig. A);
Remove the transparent pricker tip (Fig. B) and wipe it with a soft
cloth wet with the afore-mentioned solution for disinfection (Fig. C).
REMEMBER
:
While reinstalling the transparent pricker tip, check if
the depth indicator is set in the position "
↑
" (Fig. A), then set the
puncture depth appropriate for yourself.
Transparent adapter for
alternative site testing
(AST)
Release button
Lancet
ejector
Slide-on
sleeve
Expiry date
Read the instructions for use
Producer
Date of last revision
Not intended for re-sterilisation
Catalogue number
Lot number
Sterile product
For single use
Number of pieces in set
Do not use if the package is damaged
Warning
Date of production
Diagnosis S.A.
ul. Gen. W. Andersa 38A
15-113 Białystok, POLAND
Marking for OPTILET mini
Marking for OPTILETS
5221
0197
Rev. 2018.01.03 /EN