Step by step instructions
HB 142 / HB 342 / ERY 142 / HCT 142
Diaglobal GmbH · Köpenicker Straße 325 · 12555 Berlin · +49 (0)30 6576 2597 · [email protected] · www.diaglobal.de 2021-01
1. Insert capillary with 10 µL
blood sample into cuvette
2. Eject blood several times
with micropipetter into
cuvette
3. Screw cap on
Turn cuvette upside down
several times
Wait 3 minutes
HB 342: wait 30 seconds
5. Zero point adjustment:
Insert an unprocessed
cuvette from package into
photometer
Zero point is stored in
memory
4. Switch photometer on with
ON/ENTER key
Wait for device check,
confirm with ON/ENTER
Select the required test,
confirm with ON/ENTER
In regard to series
measurement:
After zero point setting any
number
of
additional samples
can be measured
ON
ENTER
ON
ENTER
HB-SLS
Insert
blank
Zero HB-SLS
0
7. Insert cuvette with blood
sample (Fig. 3) into photometer
Read measured value
ON
ENTER
HB-SLS
Sample 01
12.6 g/dL
6. Remove cuvette after
signal tone
ON
ENTER
HB-SLS
Sample 01
insert