Standard protocols for DNA shearing
Standard operating conditions
Sample volume:
100 µl
Tubes:
0.65 ml Bioruptor
®
Microtubes (Cat. No. WA-005-0500)
Tube holder:
0.65 ml tube holder for Bioruptor
®
Pico (Cat. No. B01200042) for 12 x 0.65 ml tubes
Sonication buffer:
TE (10 mM Tris, 1mM EDTA, pH 7.5 - 8.0)
DNA concentration:
1-20 ng/µl (10 ng/µl recommended)
Samples are vortexed (5-10 sec) and centrifuged (10 sec) before shearing.
For optimal results samples should be stored on ice during 5-10 minutes prior to sonication.
Temperature:
4°C –
Water Cooler (Cat. No. BioAcc-Cool) & Single Cycle Valve (Cat. No. B02020004)
Sonication cycle & total sonication time:
varies depending on desired DNA size (see table)
Note:
Recommended protocols are subject to change without notice. Additional protocols are available on demand.
125 bp
135 bp
183 bp
200 bp
234 bp
319 bp
476 bp
696 bp
848 bp
1036 bp
1230 bp
Programmable DNA size distributions,
excellent reproducibility, and high dsDNA
yields with Bioruptor
®
Pico
Figure shows different DNA size distributions
of sheared genomic DNA produced by varying
the duration of sonication. The different
curves depict a specific Bioruptor
®
Pico
Next-Generation run, optimized to produce
specific mean sizes and size ranges for Next-
Generation sequencing.
All samples were analyzed on Bioanalyzer 2100
using DNA High Sensitivity chip.