Specifications
Display Backlight
Replacing the Fuses
Features
Enclosure
Double molded, waterproof
Shock (Drop Test)
2 meters
Diode Test
Test current of 0.9mA maximum, open
circuit voltage 2.8V DC typical
Continuity Check
Audible signal will sound if the resistance
is less than 100Ω (approx.), test current
<0.35mA
Temperature Sensor
Requires type K thermocouple
Input Impedance
>10MΩ VDC & >10MΩ VAC
AC Response
True rms
AC True RMS:
The term stands for “Root-Mean-
Square,” which represents the method
of calculation of the voltage or current
value. Average responding multimeters are
calibrated to read correctly only on sine
waves and they will read inaccurately on
non-sine wave or distorted signals. True
rms meters read accurately on either type
of signal.
ACV Bandwidth
40Hz to 1000Hz
Crest Factor
≤3 at full scale up to 500V, decreasing
linearly to ≤1.5 at 1000V
Display
6,000 count backlit liquid crystal with
bargraph
Overrange indication
“OL” is displayed
Auto Power Off
15 minutes (approximately) with disable
feature