![DH Instruments PK-7000-MPC1-D Installation Instructions Download Page 2](http://html1.mh-extra.com/html/dh-instruments/pk-7000-mpc1-d/pk-7000-mpc1-d_installation-instructions_2491915002.webp)
PK-7000-MPC1-D Interconnections Kit
PG7000, RPM, PPC and MPC are trademarks, registered and otherwise, of
DH Instruments, Inc.
Document 560039
010509
DH Instruments, Inc.
4765 East Beautiful Lane
Phoenix AZ 85044-5318
USA
Tel 602.431.9100
Fax 602.431.9559
www.dhinstruments.com
q
Screw 1/8 in. tube x 1/8 in. NPT male adaptor into MPC1 TEST port using Teflon tape as a seal.
r
a To use quick connect stand for TEST connection: Install reducing union into quick connecting head.
Mount quick connect stand on bench using mounting holes provided.
r
b To use 1/4 in. NPT female adaptor for TEST connection: Install 1/8 in. tube x 1/4 in. NPT female
adaptor onto ends of tube from MPC1-D TEST port (Step
u
below).
s
Screw 1/8 in. tube x 1/8 in. NPT male adaptor into MPC1 REF port using Teflon tape as a seal.
t
Connect PG7000 or RPM3 to MPC1-D REF port by cutting a piece of 1/8 in. SS tubing to desired
length and swage together.
u
Connect quick connect stand or 1/8 in. tube x 1/4 in. NPT female adaptor to MPC1-D TEST port by
cutting a piece of 1/8 in. SS tubing to desired length and swage together.
•
1/8 in. SS tubing can be bent by hand to desired shape but use caution not to kink the tube.