MFC SWITCHBOX USER’S MANUAL 

© 1997-2008 DH Instruments, a Fluke Company 

Page 8

 

 

When the selected channel is associated with a ‘measure’ profile type, the control signal is sent to the 
device on channel 1.  This setpoint must be set in voltage and will set flows according to the range 
specified in the profile for the device connected to channel 1.  The measurements will be taken from the 
selected channel using the profile associated with the selected channel to define the flow value. 

 

 

If the profile for channel 1 indicates a current based MFC, then attempting to select a channel with a 
‘measure’ profile type will result in the error message <Voltage mode needed on channel 1 control>. 

 

 

Pressing the [ð] arrow key any time the MFC control function is active causes an instant display of 
the current MFC profile summary allowing a quick check of the characteristics of the currently active 
MFC profile.  Pressing [ESCAPE] returns from MFC profile screen to the last run screen. 

3.3

 

MFC PROFILES 

MFC profiles are used by molbox1 or MFC-CB to configure the MFC function.  MFC profiles define the 
analog signal range of the MFC or MFM and the relationship between the analog signal range and the 
flow range of the MFC or MFM.  This information allows molbox1 or MFC-CB to operate with the correct 
signal for the selected MFC and to convert the MFC or MFM analog signals to the corresponding flow or 
% FS values, when desired. 

MFC profiles are set up using 

[SETUP]

<1MFC>

.  MFC profiles are stored by molbox1 and MFC-CB 

under profile numbers.  This is the same number that will be used to select the MFC profile when the 
MFC function is activated using the MFC function key.  Up to 26 MFC profiles can be defined and stored. 

MFC profiles are designed to allow the user to set up the profiles of the MFCs that are typically used with 
so that they can be conveniently recalled from a lookup table when needed. 

 

 

MFC profiles #1 and #2 are default profiles for working with the MFC function directly in analog units 
without an MFC flow range or units.  Profile #1 is 0 to 5 V, Profile #2 is 4 to 20 mA.  When the MFC 
function is active with Profile #1 or #2 selected, the MFC units of measure will always be V or mA.   
The UNIT function selects the molbox1 flow measurement unit. 

3.4

 

UNITS OF MEASURE 

When using the MFC Switchbox with molbox1 

When the MFC function is active, the molbox1 flow units of measure will always be the units of measure 
specified in the active MFC profile. 

When the MFC function is active, the 

[UNIT]

 function key is for selection of the units of measure for the MFC 

setpoint and output.  The choices available are 

<V>

 or 

<mA>

 (depending on the active MFC profile's analog 

range), % FS or flow.  The % FS selection will cause MFC setpoint and output to be displayed in % FS of the 
MFC which molbox1 will calculate using the definition of analog range and flow range contained in the MFC 
profile.  The 

<3flow>

 selection will cause MFC setpoint and output to be displayed in the flow units of the 

MFC using the definition of analog range, flow range and flow units contained in the MFC profile. 

When using the MFC Switchbox with MFC-CB 

Units of measure are set using 

[UNIT]

.  If MFC profile #1 or #2 is active, the only units available are V, 

mA or % FS.  If an MFC profile that specifies a flow range (any profile >#2) is selected, the display unit of 
measure may be a flow unit.  See the MFC-CB Operation and Maintenance Manual for additional information. 

Summary of Contents for MFC SWITCHBOX

Page 1: ...1997 2008 DH Instruments a Fluke Company MFC Switchbox User s Manual...

Page 2: ...ever no warranty expressed or implied is provided DH Instruments disclaims any responsibility or liability for any direct or indirect damages resulting from the use of the information in this manual o...

Page 3: ...SPECIFICATIONS 3 1 4 2 MFC CONTROL FUNCTION SPECIFICATIONS 3 1 4 2 1 MOLBOX1 OR MFC CB MFC PORT 3 1 4 2 2 MFC PORT 1 DEVICE 1 3 1 4 2 3 MFC MFM PORT 2 5 DEVICE 2 DEVICE 5 4 1 4 2 4 MFC SWITCHBOX CONTR...

Page 4: ...2008 DH Instruments a Fluke Company Page II 6 TROUBLESHOOTING 13 6 1 SYMPTOMS CAUSES SOLUTIONS 13 7 MENU TREE 15 7 1 OVERVIEW 15 7 2 MOLBOX1 OR MFC CB MFC FUNCTION KEY 15 7 3 MOLBOX1 OR MFC CB SETUP...

Page 5: ...Page III 1997 2008 DH Instruments a Fluke Company T TA AB BL LE ES S Table 1 Troubleshooting Checklist 13 Table 2 DHI Authorized Service Providers 17 F FI IG GU UR RE ES S Figure 1 Front Panel 2 Figu...

Page 6: ...MA AN NU UA AL L Manual Conventions CAUTION is used in throughout the manual to identify user warnings and cautions NOTE is used throughout the manual to identify operating and applications advice and...

Page 7: ...uous power on all five devices under test DUTs The MFC Switchbox provides up to 55 watts of power This allows it to provide 260 mA per MFC when fully loaded MFCs typically require a maximum of 200 mA...

Page 8: ...2 Rear Panel 1 3 1 3 SUBASSEMBLY DESCRIPTION 1 3 1 POWER SUPPLY Powers the electronics inside the MFC Switchbox and provides continuous power to up to five connected DUTs The switching power supply pr...

Page 9: ...CB Operation and Maintenance Manual Current based MFCs can only be used when connected to device 1 of the MFC Switchbox It is not possible to switch between multiple current based MFCs The Valve Test...

Page 10: ...l Signals 16 15 VDC 15 VDC Power Supply Common Setpoint Voltage 0 to 5 VDC Setpoint Common Setpoint Sense Common MFC Measure 0 to 5 VDC MFC Measure Common Setpoint Sense 0 to 5 VDC Case Ground Earth G...

Page 11: ...INITIAL SETUP 2 3 1 POWER CONNECTION Connect the MFC Switchbox to a power source using the power cable supplied Power requirements are 85 to 264 VAC 47 to 440 Hz 55 VA maximum consumption 2 3 2 CONNE...

Page 12: ...MFC SWITCHBOX USER S MANUAL 1997 2008 DH Instruments a Fluke Company Page 6 N NO OT TE ES S...

Page 13: ...g the setup menu the MFC channel can be selected by pressing the MFC function key on the molbox1 or MFC CB front panel When the MFC function key is first pressed the screen prompts the operator to sel...

Page 14: ...unction is activated using the MFC function key Up to 26 MFC profiles can be defined and stored MFC profiles are designed to allow the user to set up the profiles of the MFCs that are typically used w...

Page 15: ...ts the MFC channel to use for optional MFC Switchbox mode The MFC mode v Voltage control and measure mode The selected channel will be used for voltage setting and measuring mA Current loop control an...

Page 16: ...MFC SWITCHBOX USER S MANUAL 1997 2008 DH Instruments a Fluke Company Page 10 N NO OT TE ES S...

Page 17: ...USTMENTS Page 11 1997 2008 DH Instruments a Fluke Company 5 5 M MA AI IN NT TE EN NA AN NC CE E A AN ND D A AD DJ JU US ST TM ME EN NT TS S 5 1 OVERVIEW There are no user maintainable parts or systems...

Page 18: ...MFC SWITCHBOX USER S MANUAL 1997 2008 DH Instruments a Fluke Company Page 12 N NO OT TE ES S...

Page 19: ...d The MFC Switchbox power is not on Power on the MFC Switchbox MFC connected to molbox1 doesn t appear to be responding MFC option not implemented Check to see if this option was ordered and installed...

Page 20: ...MFC SWITCHBOX USER S MANUAL 1997 2008 DH Instruments a Fluke Company Page 14 N NO OT TE ES S...

Page 21: ...s a Fluke Company 7 7 M ME EN NU U T TR RE EE E 7 1 OVERVIEW The following is the portion of the molbox1 or MFC CB internal menu that pertains to use of the MFC Switchbox 7 2 MOLBOX1 OR MFC CB MFC FUN...

Page 22: ...MFC SWITCHBOX USER S MANUAL 1997 2008 DH Instruments a Fluke Company Page 16 N NO OT TE ES S...

Page 23: ...ill even if DHI has been advised of the possibility thereof and regardless of whether such products are used individually or as components in other products Items returned to DHI under warranty claim...

Page 24: ...MFC SWITCHBOX USER S MANUAL 1997 2008 DH Instruments a Fluke Company Page 18 N NO OT TE ES S...

Reviews: