background image

 

molbloc, molbloc-S, molbox, molstic and molstic-s are trademarks, registered and otherwise, of 

DH Instruments, Inc. 

VCR is a registered trademark of the Swagelok Company.

 

Document 560056 

041215 

DH Instruments, Inc. 

4765 East Beautiful Lane 

Phoenix  AZ  85044-5318 

USA 

 

 

Tel 602.431.9100

Fax 602.431.9559

[email protected]

www.dhinstruments.com

 

  Align the filter downstream fitting (4) in the VCR nut 

(5) and thread the nut onto the fitting so it is hand-
tight.  The filter should be installed with the printed 
arrow pointing in the direction of the gas flow. 

  Using two wrenches, tighten the filter’s male VCR 

connector (4) into the female nut (5) approximately 
1/2 turn beyond hand-tight. 

  Thread the 8VCR female coupling (2) onto the filter 

upstream connector (3).  Place a new 8VCR o-ring 
securely against the sealing surface of the filter 
upstream fitting inside the 8VCR female 

 

coupling. 

  Thread the VCR x tube adaptor fitting (1) from the 

original molstic inlet into the 8VCR female coupling 
(2).  The coupling should be threaded onto both 
male fittings so that the o-ring is centered in the 
coupling. 

  Using two wrenches, tighten the two male nuts from 

the tube adaptor and filter upstream fitting against 
each other in the coupling.  The o-ring connection 
should be tightened only 1/2 turn past finger-tight. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Over-tightening VCR connections will damage the sealing 
beads and/or o-rings and possibly cause system leakage.  Do 
not re-use metallic VCR gaskets.

 

 

The following parts are included in the shipment: 

DESCRIPTION 

QTY 

PART NO. 

Filter, SS, 5 micron, 8VCR  

103129 

Adaptor, 8VCR female coupling 

103606 

Gasket, 8VCR 

102923 

O-ring, Viton, 8VCR 

102912 

 
 
 
 

 