Page 37 of 53
8000-INS-UMENG Rev. 4
15.25 Voice Controls
Some controls can be operated via verbal commands from the user. Voice control can be
enabled/disabled through the
Preferences
→
System
menu. Voice control can be
configured to begin upon pressing the “Enter” key or by a verbal command of “Hey
Flex”. Voice control sensitivity can be adjusted in this menu as well. The following
verbal commands are recognized by the Scanmate software while performing a B-Scan.
User Verbal
Commands
Software Audible
Response
Software Command
Executed
“Hey Flex”
Tone: Double-Beep
Voice control activated. The software is
ready to receive verbal commands.
“Start Scan”
Verbal: “Starting Scan” The B-scan will start.
“Stop Scan” *
Verbal: “Scan Stopped” The B-scan will stop.
“Change Eye”
Verbal: “Eye changed to
OD/OS.”
The eye being scanned is changed. The
audible response will state if OD or OS is
being scanned.
“Zoom In”
N/A
Zoom will enlarge the image or video.
“Zoom Out”
N/A
Zoom will reduce the image or video.
“Show Ruler”
N/A
The ruler overlay is shown.
“Hide Ruler”
N/A
The ruler overlay is hidden.
“Save Image”
Verbal: “Image Saved” The currently displayed frame is saved.
“Save Video”
Verbal: “Video Saved” The currently displayed video is saved.
“Open File”
N/A
A window is opened showing
B-Scan records for the currently loaded
patient.
“Yes” *
N/A
Selects ‘Yes’ for a dialog box prompts.
“No” *
N/A
Selects ‘No’ for a dialog box prompts.
* Indicates a command that DOES NOT require “Hey Flex” or an “Enter” key press to be
initiated.