8 Technical data
Tube head assembly
Tube head assembly type
THA-I
Tube type
D-0711SB or Kailong KL 21 SB or equivalent
Max. tube voltage
60 or 70 kV
Max. tube current
7 mA
Max. electric output
490 W (70kV; 7mA)
Reference axis
Runs axially with the cone
Target angle
16 degrees
Focal spot
0.7 mm (according to IEC 60336/2005)
Nominal anode input power
940 W
Max. symmetrical radiation field
Ø 60 mm at a 200 mm focal length
Total filtration
2,0 mm Al (70 kV)
Inherent filtration
1,0 mm Al (70 kV)
Fixed additional filtration
1,0 mm Al (70 kV)
Max. anode heat content
7 kJ
Maximum X-ray tube assembly heat
content
140 kJ
Maximum continuous heat dissipation of
the X-ray tube head assembly
19 W
Dental care unit mount model device
Nominal mains voltage
115 VAC +/- 10%
230 VAC +/- 10%
Input power frequency
60 Hz
50 Hz
Nominal current
7.63 A
3.55 A
Mains fuse, slow blow
6.25 AT
Apparent resistance of supply mains
0.68 ohm
Power consumption
816 VA / 230 VAC
877 VA / 115 VAC
US/Canada mains connector type
115 V / NEMA 6-15P or similar
Power supply cords type
H05VV5-F / AWG 14 (UL 2587)
FOCUS
™
29