Table 5. Hypoglycemic Alert Evaluation
Hypoglycemic
Alert Level
(mg/dL)
Study
1
True
Alert
Rate
False
Alert
Rate
Hypoglycemia
Detection Rate
Hypoglycemia
Missed Detection
Rate
55
Original
50%
50%
71%
29%
Software 505
71%
29%
68%
32%
60
Original
64%
36%
75%
25%
Software 505
85%
15%
83%
17%
70
Original
79%
21%
83%
17%
Software 505
92%
8%
91%
9%
80
Original
87%
13%
86%
14%
Software 505
95%
5%
90%
10%
90
Original
90%
10%
89%
11%
Software 505
96%
4%
94%
6%
1
Both sets of study data are presented and are labeled as
Original
(SW10050) or
Software 505
(SW10505).