![Dewetron HSI-LV Technical Reference Manual Download Page 9](http://html1.mh-extra.com/html/dewetron/hsi-lv/hsi-lv_technical-reference-manual_2489963009.webp)
9
HSI-LV
High speed isolated voltage amplifier
Bandwidth:
2 Mhz
Input ranges:
12 ranges (10 mV to 50 V)
Input type:
AC and DC coupling software selectable
TEDS:
Supports electronic data sheet sensors
Signal connection:
HSI-LV-B:
Banana plugs
HSI-LV-BNC:
BNC connector
HSI-LV-D:
9-pin D-SUB connector
HSI-LV-LEMO:
7-pin LEMO connector
Module specifications
HSI-LV
Input ranges
10 mV, 20 mV, 50 mV, 100 mV, 200 mV, 500 mV, 1 V, 2.5 V, 5 V, 10 V, 25 V, 50 V
Button selectable ranges
10 mV, 50 mV, 200 mV, 1 V, 5 V, 10 V, 50 V
Rated input voltage
33 V
RMS
, 46.7 V
PEAK
, 70 V
DC
according to EN-61010-1 and EN-61010-2-30
1 year accuracy
1)
Range
Signal frequency
Accuracy
Bipolar
10 mV to 100 mV
DC
±0.02 % of reading ±60 µV
2.5 V
DC
±0.02 % of reading ±0.1 % of range
200 mV to 50 V
DC
±0.02 % of reading ±0.05 % of range
10 mV to 100 mV
0.1 Hz to 5 kHz
±0.1 % of reading ±30 µV
>5 kHz to 50 kHz
±0.4 % of reading ±30 µV
>50 kHz to 100 kHz
±(0.016*f) % of reading ±0.1 % of range
>100 kHz to 1 MHz
±(0.010*f) % of reading ±1 % of range
>1 MHz to 2 MHz
±(0.014*f) % of reading ±3 % of range
200 mV to 50 V
0.1 Hz to 500 Hz
±0.05 % of reading ±0.01 % of range
>500 Hz to 5 kHz
±0.1 % of reading ±0.05 % of range
>5 kHz to 50 kHz
±0.4 % of reading ±0.05 % of range
>50 kHz to 100 kHz
±(0.016*f) % of reading ±0.1 % of range
>100 kHz to 1 MHz
±(0.010*f) % of reading ±1 % of range
>1 MHz to 2 MHz
±(0.014*f) % of reading ±3 % of range
Unipolar
10 mV to 100 mV
DC
±0.02 % of reading ±60 µV
200 mV to 50 V
DC
±0.02 % of reading ±0.08 % of range
Input coupling
DC or AC software selectable (1.5 Hz standard, custom on request down to 0.01 Hz)
Gain linearity
Typically 0.01 %; max. 0.04 % of full scale
Gain drift range
Typically 10 ppm/°C (max. 30 ppm/°C)
Off set drift
10 mV to 200 mV:
Typically 3 µV/°C
500 mV to 50 V:
Typically 10 ppm of range/°C
Long term stability
100 ppm/sqrt (1000 hrs)
Input resistance
1 MOhm
Bandwidth (-3 dB)
2 MHz
Signal delay @ full bandwidth
approx. 405 ns
Filter selection
Push button or software
Filter
100 Hz, 300 Hz, 1 kHz, 3 kHz, 10 kHz, 30 kHz, 100 kHz, 300 kHz, 1 MHz, 2 MHz
2)
Filter type
Bessel or Butterworth 40 dB/dec
Filter characteristics
100 Hz to 1 MHz
2 MHz
Butterworth or Bessel 40 dB/dec (2
nd
order; ±1.5 dB @ f
0
)
Butterworth 60 dB/dec (3
rd
order; 0 to -3 dB @ 2 MHz)
Typical SFDR and SNR:
20 mV
1 V
50 V
10 kHz bandwidth
100 kHz bandwidth
1 MHz bandwidth
2 MHz bandwidth
SFDR SNR
SFDR SNR
SFDR SNR
SFDR SNR
88 dB 78 dB
88 dB 71 dB 77 dB 60 dB
76 dB 56 dB
110 dB 98 dB
110 dB 95 dB 93 dB 82 dB
84 dB 75 dB
110 dB 98 dB
110 dB 95 dB 94 dB 82 dB
85 dB 75 dB
Typical CMRR
50 Hz
1 kHz
10 kHz
100 kHz
10 mV to 1 V range:
2.5 V to 50 V range:
130 dB
100 dB
120 dB
60 dB
95 dB
40 dB
75 dB
20 dB
Input overvoltage protection
350 V
DC
Isolation voltage
1 kV
RMS
3)
Sensor supply
±9 V (±1 %), 12 V (±5 %), 200 mA resettable fuse protected
4)
Output voltage
±5 V
Output resistance
10 Ohm
Maximum output current
5 mA
Output protection
Short to ground for 10 sec.
Power On default settings
Software programable
Power supply
±9 V
DC
±1 %
Power consumption
1.1 W without sensor supply
Special functions
Integrated temperature sensor
RS-485 interface
Yes
TEDS
Hardware support for TEDS (Transducer Electronic Data Sheet)
Supported TEDS chips
DS2406, DS2430A, DS2432, DS2433, DS2431
Supported MSI
MSI-V-ACC, MSI-V-RTD
1)
Conditions for accuracy: Module temperature is calibration temperature ±5 °C; humidity is 30 to 90 RH.
AC accuracy: the highest filter (2 MHz) has to be activated. f = signal frequency in kHz.
For the 2 year accuracy multiply all % of range and % of reading values by 1.5.
2)
2 MHz filter: exclusively for Butterworth 60 dB/decade - refer to filter specifications. Please consider possible bandwidth limitation of further components in the
measuring chain, e.g. A/D card or signal conditioning mainframe.
3)
Although the rated input voltage is 33 V
RMS
, 46.7 V
PEAK
or 70 V
DC
according to EN-61010-1 and EN-61010-2-30, the galvanic isolation has been tested with
1 kV
RMS
for 1 min.
4)
Overall current should not exceed DEWE-30-xx maximum power.
continued on next page
nical Refe
450134 • DAQP/H
SI
/PAD Modules • Tech
rence Manual • Prin
t
ing version 2.3.4 • November 18, 2020
Summary of Contents for HSI-LV
Page 16: ...C2 Notes...