10
HSI-HV
IN+
IN-
Safety Banana plugs
red
black
INPUT
CONNECTOR
RS-485(A)
RS-485(B)
+9V
+12 V
reserved
reserved
-9 V
±5 V out
GND
2
3
5
6
7
8
9
1
4
OUTPUT
CONNECTOR
10k
10k
Programmable
Amplifier
Isolation
2nd order
lowpass filter
(BE or BU)
100 Hz to 1 MHz
5M 0.1 %
5M 0.1 %
Common Mode
Rejection Trim
HSI-HV module
9-pin DSUB
3rd order
Butterworth filter
2 MHz
or off
GND
ISO
Front panel control
LED indication:
The HSI-HV series module has a set of 8 LEDs showing the current input range (constant active)
and filter range (flashing) setting.
Filter
Range
Range
Filter
20 Hz
1400 V
100 V
10
5
10 Hz
800 V
50 V
10
4
3 Hz
400 V
20 V
10
3
1
Hz
200
V
10
2
Push button operation:
Select range:
Push the
RANGE
button several times shortly until the LED displays the
desired input range.
Select filter:
Push the
FILTER
button once - the LEDs will flash for approx. 3 seconds and
display the current filter setting. Push the
FILTER
button within the three seconds
several times until the flashing LED displays the desired filter range.
Power On Default function
You can store the actual settings of the module in the internal EE-Prom memory. Once the module restarts,
it comes up automatically with these setting. This is important for stand alone applications and for fail save
reasons. If the function is deactivated the module automatically remembers the last pushbutton selected
range and filter.
Block diagram
The basic block diagram of the HSI-HV gives an idea of the internal structure.
Summary of Contents for HSI-HV
Page 14: ...C2 Notes...