background image

9

DAQP-ACC-A

High dynamic signal amplifier

Supported sensors: 

IEPE

®

 sensors

Input sensitivity: 

±5 V, ±1.66 V, ±500 mV, ±166 mV, ±50 mV 

(Gain 1, 3, 10, 30 and 100), 

push button or software selection

Bandwidth: 

300 kHz

Sensor connection: 

BNC connector

Module specifications

IEPE

®

 is a trademark of PCB Piezotronics, Inc.

DAQP-ACC-A

Ranges

±5 V, ±1.66 V, ±500 mV, ±166 mV, ±50 mV

Gain

1, 3, 10, 30, 100

Range / gain selection

Push button or software

Gain error

0.5 %

Sensor types

IEPE

®

 sensors only

Sensor excitation

4 or 8 mA (software selection), 10 %, up to 28 V

DC

Input impedance

5 or 7 MOhm (depending on time constant), in parallel with 1.2 nF

Input voltage range

  Voltage < 4 V

  Voltage > 19 V

4 to 19 V

‚Shortcut‘ detection

‚No sensor‘ detection

Input protection

  IN+

  IN- (shield)

max. -10 to 28 V

max. 20 mA

Bandwidth (-3 dB)

From selected highpass fi lter to 300 kHz (+2 to -5 dB @ fg)

Filters (highpass)

  0.5 Hz fi lter

  5 Hz fi lter

0.5 Hz and 5 Hz (software selection)

0.32 s time constant

0.032 s time constant

Filters (lowpass)

1 kHz, 10 kHz, 100 kHz, 300 kHz

other fi lter steps available as an option on request

Filter selection

Push button or software

Filter characteristics

  up to 100 kHz

  300 kHz

Butterworth

100 dB / decade (30 dB / octave)

80 dB / decade (24 dB / octave)

Typical SFDR and SNR

  5000 mV

  500 mV

  50 mV

300 kHz bandwidth 

         100 kHz  bandwidth 

      10 kHz  bandwidth

SFDR           SNR 

         SFDR 

        SNR 

      SFDR 

   SNR

92 dB           90 dB 

         100 dB          91 dB 

      103 dB 

   93 dB

88 dB           85 dB  

         100 dB          89 dB 

      100 dB 

   92 dB

71 dB           68 dB 

          90 dB 

        73 dB 

       82 dB 

   80 dB

Output voltage

±5 V

Output resistance

< 10 Ohm

Output current

Max. 5 mA

Output protection

Continuous short to ground

RS-485 interface

Yes

Power supply voltage

±9 V

DC

 (±10 %)

Power consumption

Typical 0.8 to 1.0 W (depending on sensor)

450134 • DAQP/H

SI

/PAD Modules • Technical Reference Manual • Prin

t

ing version 2.3.4 • November 18, 2020

Summary of Contents for DAQP-ACC-A

Page 1: ...l supply you with accurate complete and reproducible measurement results for your decision making Look forward to the easy handling and the flexible and modular use of your DEWETRON product and draw u...

Page 2: ...protected by copyright All rights are reserved Reproduction adaptation or translation without prior written permission is prohibited except as allowed under the copyright laws All trademarks and regis...

Page 3: ...se of the information in this document Warranty Information A copy of the specific warranty terms applicable to your DEWETRON product and replacement parts can be obtained from your local sales and se...

Page 4: ...wing general safety precautions must be observed during all phases of operation service and repair of this product Failure to comply with these precautions or with specific warnings elsewhere in this...

Page 5: ...ained Keep away from live circuits Operating personnel must not remove equipment covers or shields Procedures involving the removal of covers or shields are for use by service trained personnel only U...

Page 6: ...com The telephone hotline is available Monday to Friday between 08 00 and 12 00 CET GMT 1 00 and Monday to Thursday between 13 00 and 17 00 CET For the Americas please contact DEWETRON Inc 2850 South...

Page 7: ...AQ series modules available The CAL KIT contains the required cables software and instructions General module specifications Module dimensions 20 x 65 x 105 mm 0 79 x 2 56 x 4 13 in W x H x D without...

Page 8: ...e FILTER button again Approx 3 seconds after the last key activity the information will be stored the LED stops flashing and shows the input range again CAUTION Power loss during this time leaves the...

Page 9: ...lected highpass filter to 300 kHz 2 to 5 dB fg Filters highpass 0 5 Hz filter 5 Hz filter 0 5 Hz and 5 Hz software selection 0 32 s time constant 0 032 s time constant Filters lowpass 1 kHz 10 kHz 100...

Page 10: ...he GAIN button several times until the LED displays the desired input range Filter button Push the FILTER button once the LEDs will flash for approx 3 seconds and display the current filter setting Pu...

Page 11: ...ber States relating to electrical equipment designed for use within certain voltage limits amended by the directive 93 68 EEC 89 336 EEC Directive on the approximation of the laws of the Member States...

Page 12: ...C2 Notes...